Flow Rate

Flow Rate Edit

Logging Parameters

Thermo Fisher Scientific
The Flow Rate screen displays the set flow rate and the corresponding
cyclone mass median diameter 50% cut point (D50).
From the Configure menu, press the
key to the Flow Rate screen.
FLOW RATE
1.19 LPM
The Flow Rate Edit screen allows the user to edit the volumetric flow rate.
The flow range is 1–3.5 L/min and if a cyclone has already been chosen,
the corresponding 50% aerodynamic cut point (D50 in micrometers) will
change simultaneously. Use the ▲ and ▼ keys to increase or decrease the
flow rate in increments of 0.01 LPM. The asterisk (*) appears to the left of
the field being edited.
From the Flow Rate screen, press the
Press the
key to save the flow rate and return to the Flow Rate
ENTER
screen.
FLOW RATE
*1.19 LPM
The Logging Parameters display screen indicates on the first line whether
data logging is enabled or disabled. The second line of this screen displays
the logging period for 3 seconds, the site label for 3 seconds, and the tag
number for 3 seconds in sequence.
From the Configure menu, press the
key to the Logging screen.
LOGGING ENABLED
1:00:00 h:m:s
LOGGING ENABLED
Factory default
ENTER
(D50)
10.00
ENTER
(D50)
10.00
ENTER
MIE ADR-1500 Instruction Manual
Operation
key and scroll the ▲ or ▼
key and edit the flow rate.
key and scroll the ▲ or ▼
3-15