Flow Rate
Flow Rate Edit
Logging Parameters
Thermo Fisher Scientific
The Flow Rate screen displays the set flow rate and the corresponding
cyclone mass median diameter 50% cut point (D50).
From the Configure menu, press the
●
key to the Flow Rate screen.
FLOW RATE
1.19 LPM
The Flow Rate Edit screen allows the user to edit the volumetric flow rate.
The flow range is 1–3.5 L/min and if a cyclone has already been chosen,
the corresponding 50% aerodynamic cut point (D50 in micrometers) will
change simultaneously. Use the ▲ and ▼ keys to increase or decrease the
flow rate in increments of 0.01 LPM. The asterisk (*) appears to the left of
the field being edited.
From the Flow Rate screen, press the
●
Press the
key to save the flow rate and return to the Flow Rate
ENTER
●
screen.
FLOW RATE
*1.19 LPM
The Logging Parameters display screen indicates on the first line whether
data logging is enabled or disabled. The second line of this screen displays
the logging period for 3 seconds, the site label for 3 seconds, and the tag
number for 3 seconds in sequence.
From the Configure menu, press the
●
key to the Logging screen.
LOGGING ENABLED
1:00:00 h:m:s
LOGGING ENABLED
Factory default
ENTER
(D50)
10.00
ENTER
(D50)
10.00
ENTER
MIE ADR-1500 Instruction Manual
Operation
key and scroll the ▲ or ▼
key and edit the flow rate.
key and scroll the ▲ or ▼
3-15