Operation
Operate Menu
Start A Run
Concentration/Scattering
3-6
MIE ADR-1500 Instruction Manual
The Operate Menu is used to access the list of displays allowing the user to
access the operating modes and run-time parameters of the ADR-1500
including "Start A Run", "Delayed Start", "Zero", "Logging Parameters"
and "Battery/Memory Status".
OPERATE
The Start A Run menu allows the user to begin monitoring/sampling. The
run will be logged only if the asterisk (*) appears at the end of the first
display line (see below). If the ADR-1500 is ready to begin a monitoring
run, press
and the instrument will begin operating as configured.
ENTER
Equipment Damage Verify that the exhaust port and the inlet port are not
blocked.
▲
From the Operate Menu, press the
●
key to the Start a Run screen.
Press the
key and the instrument will begin operating as
ENTER
●
configured.
START A RUN
Once a run is started, press the ▲ and ▼ keys to view a series of the
following displays.
The instrument measures concentration or scattering coefficient of ug/m
for concentration or 1/Mm for scattering coefficient. The real-time
concentration is presented on the first line and the time-weighted average is
presented on the second line. The real-time concentration is an average
taken over a time interval called "Display Time", which is programmed by
the user. The presence of the asterisk (*) in the first line indicates that data
logging is enabled.
key and scroll the ▲ or ▼
ENTER
*
3
Thermo Fisher Scientific