Pressure Offset

Pressure Offset Edit

RH Offset Edit

Thermo Fisher Scientific
The Pressure Offset screen allows the user to view and edit the pressure
offset of the barometric pressure measurement to coincide with their
traceable standards.
From the Calibrate Menu, press the
key to the Pressure Offset screen.
The Pressure Offset Edit screen allows the user to edit the pressure offset.
Use the ▲ and ▼ keys to increase or decrease the user pressure offset in
increments of 1 in millimetres of mercury. The asterisk (*) appears to the
left of the field being edited.
From the Pressure Offset screen, press the
pressure offset.
Press the
Pressure Offset screen.

RH Offset

The RH Offset screen allows the user to view and edit the RH offset of the
barometric pressure measurement to coincide with their traceable
standards.
From the Calibrate Menu, press the
key to the RH Offset screen.
The RH Offset Edit screen allows the user to edit the RH offset. Use the
▲ and ▼ keys to increase or decrease the RH offset in increments of 0.1%
RH. The asterisk (*) appears to the left of the field being edited.
PRESSURE OFFSET
0
760 mmHg
key to save the user pressure offset and return to
ENTER
PRESSURE OFFSET
*
10
750 mmHg
RH OFFSET
0.0
47.3 %
key and scroll the ▲ or ▼
ENTER
key and edit the
ENTER
key and scroll the ▲ or ▼
ENTER
MIE ADR-1500 Instruction Manual
Operation
3-25