Guidelines and Instrument Layout
2-20
MIE ADR-1500 Instruction Manual
appear on the computer screen. From the menu bar or the embedded
Settings window, the serial connection port can be selected (e.g., COM
4). Select the port to which the USB cable has connected to on you
computer using the Select Port pop-up window and click OK to
proceed. The user may now click on the Show Instrument Panel to
emulate the instrument keypad or utilize the tabs within the pDR Port
notebook display.
Most operations with the pDR Port software program are self-evidently
labeled, including fly-over dialog boxes. In addition, instructions may be
found in the On-line Help files by selecting Help and then Contents.
The following operating/logging parameters of the ADR-1500 can be
selected (edited) via the computer:
Current date (year, month and day of the month)
Current time (hour, minute and second)
Display averaging time (1 to 60 seconds, in 1-second increments)
Calibration factor (0.01 to 9.99, in 0.01 increments)
Analog output full scale concentration (0.1, 0.4, 1, 4, 10, 40, 100, or
3
400 mg/m
Analog output status (enabled or disabled)
Alarm level (0.01 to 400.0 mg/m
Alarm status (enabled or disabled)
Humidity correction (enabled or disabled)
3
3
, in 1-g/m
increments)