Maintenance and Service
Bracket Assembly and
Power Supply/Charger
Replacement
5-30
MIE ADR-1500 Instruction Manual
Use the following procedure to replace the bracket assembly and power
supply/charger
(Figure
1. Turn the instrument OFF. Remove the AC plug from the bottom of
the instrument.
2. Unfasten the two captive hardware and swing cover down from the
power/com assembly and on the center panel.
3. Disconnect the battery cable from J6 or J8 from the power
communication board.
4. Disconnect connectors from board and cable power supply. Note the
locations of the connectors to facilitate re-connection. See
a. Power supply disconnect:
4-pin power supply to cable power supply
2-pin to 24 V J1 on the power communication board
b. Charger disconnect:
3-pin charger to cable power supply
4-pin charger to J6 or J8 to power communication board
5. Unfasten four keps nut and remove brackets.
6. Replace bracket assembly and power supply charger following the
previous steps in reverse order.
5–14).
Figure
5–7.