Troubleshooting
6-4
MIE ADR-1500 Instruction Manual
Malfunction
Possible Cause
No USB
Wrong COM port (pDR Port)
communication
Disconnected cable
Defective external USB cable
Defective internal USB cable
Non-working strobe
Disconnected cable
Defective strobe
No relay (if installed)
Disconnected cable
Defective relay
Action
Verify COM port (pDR Port).
Verify cable.
Replace cable.
Send in for service.
Verify cable connection.
Send in for service.
Verify cable connection.
Replace relay.