Operation

RH Correction

RH Correction Edit

3-18
MIE ADR-1500 Instruction Manual
INLET TYPE
*RED CYCLONE
The RH Correction screen allows the user to view whether RH correction
is enabled or disabled.
From the Configure menu, press the
key to the RH Correction screen.
RH CORRECTION
Note RH Correction should be enabled when the RH is expected to
exceed 50–60%. Ambient monitoring should always use the RH
Correction as being enabled. However, precision can be affected between
collocated devices.
The RH Correction Edit screen allows the user to enable or disable RH
correction. The asterisk (*) appears to the left of the field being edited
From the RH Correction screen, press the
correction.
Press the
Correction screen.
RH CORRECTION
* ENABLED

Alarm

The Alarm screen indicates the alarm options for triggering the alarm
(Disabled, Instant, or STEL) and the value at which it is triggered. The
instant setting causes the alarm to be triggered when the displayed
concentration is higher than the instant alarm level setting. The STEL
setting causes the alarm to be triggered when the 15 minute rolling
(boxcar) average is higher than the STEL alarm level setting. Each of the
two enabled alarm options use separately stored alarm level values.
DISABLED
key to save the RH correction and return to the RH
ENTER
key and scroll the ▲ or ▼
ENTER
key to edit the RH
ENTER