Alert
Description
Whenever the ADR-1500 is shipped, care should be taken
in repackaging it with the original factory provided
packaging.
Damage could occur if not installed in a vertical, upright,
position.
Damage may occur to the instrument if the environmental
conditions exceed IP65.
Equipment damage may occur to instrument if power
inputs or fuse type exceeds specified ranges.
Equipment damage may occur if exhaust port is blocked
or port covers are not in place if unused.
Do not attach additional items to the handle for vertical
hauling as this might compromise the strength of the
handle and the ADR-1500 enclosure.
It should be noted that the ADR-1500 can be powered
from any line with a voltage between 100-240 volts A.C.,
50 to 60 Hz. No internal adjustments or selections need
to be made for power lines with voltages and frequencies
in those ranges. The internal AC-to-DC power unit
performs any adjustments automatically.
Please note that the important purpose of the HEPA filter
or filter cassette is to protect the pump.
At no time should the ADR-1500 be running without a
filter in place, otherwise serious damage to the pump
components may result.
Plugging or unplugging any external equipment (e.g.,
computer, modem, alarm circuitry, etc.) should be made
only while both the ADR-1500 and the external
equipment are shut off, in order to prevent damage or
interference due to transient electrical effects.
It is recommended to turn the instrument and computer
OFF before making a connection.
At no time should the ADR-1500 be running without a
filter in place, otherwise serious damage to the internal
components may result.
Unless a MALFUNCTION message is displayed, or other
operational problems occur, the ADR-1500 should be
returned to the factory once every year after being placed
into service for routine check out, test, cleaning and
calibration check.
Some internal components can be damaged by small
amounts of static electricity. A properly grounded
antistatic wrist strap must be worn while handling any
internal component.