Maintenance and Service
Removing the
Internal Covers
5-8
MIE ADR-1500 Instruction Manual
The internal covers can be removed to allow access to the core components
of the ADR-1500. Partial removal allows access to the power/com assembly
and center panel of the ADR-1500 and full removal allows access to
everything on the ADR-1500. Refer to the following steps when a
procedure requires partial or full removal of the internal cover
Equipment Damage Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
1. Turn the instrument OFF. Remove the AC plug from the bottom of
the instrument.
2. For access to the power/com assembly, unfasten the two captive
hardware and swing cover down. For center panel removal, unfasten
the two captive hardware and swing cover down. For full removal,
unfasten the two captive hardware and six screws.
3. Disconnect the battery cable from J6 or J8 from the power
communication board.
4. Re-install the internal cover by following the previous steps in reverse
order.
(Figure
5–2).