Start Time/Date Edit

Delayed Start Countdown

Zeroing the ADR-1500

3-10
MIE ADR-1500 Instruction Manual
START@
14:45:30
13-Nov-09
The Start Time/Date Edit screen allows the user to edit the delayed start
time and date. In the display, the
navigation through six fields within the display in the following sequence:
hours, minutes, seconds, years, months, and days. The
navigates forward while the Esc key navigates backwards. The ▲ and ▼
keys are used to increment/decrement the field values. The asterisk (*)
appears to the left of the field being edited.
From the Start@ screen, press the
next field, except if day (of month) is selected: the delayed start time
and date saved, the delayed start status is enabled, and the display
advances to Delayed Start Countdown.
START@ *14:45:30
13-Nov-09
The Delayed Start Countdown screen displays the time of day that the
ADR-1500 will start a run on the first line and the amount of time
remaining before the instrument starts this run in days, hours, minutes,
and seconds. The presence of the asterisk in the first line indicates that data
logging is enabled.
START@ *14:45:30
120Days hh:mm:ss
The Zero Instrument screen is used to initiate a zero sequence to measure
and store the optical background. The second line displays advice to the
user that a filter must be applied to the Total Inlet to provide clean
(particulate-free) air to the optics for this process.
From the Operate menu, press the
key to the Zero Instrument screen.
Note Do not zero through the red or blue sharp-cut cyclones. Only zero
through a clean Total Inlet.
and the
ENTER
ESC
key and to advance to edit the
ENTER
key and scroll the ▲ or ▼
ENTER
keys provide
key
ENTER
Thermo Fisher Scientific