Operation
Calibrate Menu
Temperature Offset
Temperature Offset Edit
3-24
MIE ADR-1500 Instruction Manual
The Calibrate Menu is used to access the list of displays allowing the user
to calibrate the sensors used to measure scattering, temperature, relative
humidity, barometric pressure, flow rate, and the time and date. The
displays include "Temp Offset", "Pressure Offset", "RH Offset", "Flow
Rate Cal", and "Cal Factor".
CALIBRATE
The Temperature Offset screen allows the user to view and edit the
temperature offset of the measurement to coincide with their traceable
standards.
From the Calibrate Menu, press the
●
key to the Temp Offset screen.
TEMP OFFSET
0.0
The Temperature Offset Edit screen allows the user to edit the temperature
offset. Use the ▲ and ▼ keys to increase or decrease the user temperature
offset in increments of 0.1 in degrees Celsius. The asterisk (*) appears to
the left of the field being edited.
From the Temp Offset screen, press the
●
temperature offset.
Press the
key to save the temperature offset and return to Temp
ENTER
●
Offset screen.
TEMP OFFSET
* 0.1
Please note that a precise and accurate calibration of the RH-sensor is
critical to maintaining precision between collocated devices when using RH
Correction enabled.
ENTER
22.3 C
ENTER
22.2 C
key and scroll the ▲ or ▼
key and edit the
Thermo Fisher Scientific