Operation

Configure Menu

Display Average
Display Average Time
3-14
MIE ADR-1500 Instruction Manual
The Configure Menu is used to access the list of displays allowing the user
to configure how the instrument measures and logs data including "Display
Avg Time", "Flow Rate", "Logging Parameters", "Inlet Type", "RH
Correction", "Alarm", "Units", "Analog Output", "LCD Backlight", LCD
Contrast" and "Time/Date".
CONFIGURE
The Display Average Time screen offers the user to adjust the running
(boxcar) average time of the displayed concentration (or scattering
Time
coefficient) between 1 and 60 seconds. The average concentration displayed
is the value scaled to the analog output and offered as an alarm criterion.
From the Configure menu, press the
key to the Display Avg Time screen.
DISPLAY AVG TIME
The Display Average Time Edit screen allows the user to change the
running average displayed on the screen between 1 and 60 seconds. This
Edit
also doubles as the running average that is provided as an analog output
and also as one of the values that triggers the alarm. Use the ▲ and ▼ keys
to increase or decrease the display average time by seconds. The asterisk (*)
appears to the left of the field being edited.
From the Display Avg Time screen, press the
display averaging time.
Press the
Display Avg Time screen.
DISPLAY AVG TIME
*
10 seconds
key to save the display averaging time and return to
ENTER
11 seconds
key and scroll the ▲ or ▼
ENTER
key and edit the
ENTER