Serial Commands
B-4
MIE ADR-1500 Instruction Manual
SITE [site #]
Returns, or programs, the description of site number sitenum. Sites are
numbered 1–50.
SD [Status]
Returns, or programs, the streaming data status.
Status = {"enabled" or "on", "disabled" or "off"}
TAG [Tag #] [status]
Returns, or programs, the tag number to be used when next time logging is
to start. And enable or disable logging.
Tag # = {00, 1, 2, ..., 99}
Status = {"enabled" or "on", "disabled" or "off"}
TAGDUMP [Tag #]
Returns the complete data logged under the file tag number.
Tag # = {0, 1, 2, ..., 99}
TAGS [Tag #], [Tag #], ...
Get a list of the tag files currently in the device.
TEMP [temp] [RH] [Pa]
Get current temperature, RH and atmospheric pressure.
temp is in degrees Celsius
RH is Relative Humidity in percent
Pa is barometric pressure in mmHG
TEMPUNITS [T]
Responds with "TEMPUNITS C", degrees Celsius are the only allowed.
TIME [seconds] [minutes] [hours]
Returns, or programs, the RTC's current time.
UNITS [integer] the current measurement units:
Units =
0 = μg/m
3
(Mass concentration)
1 = 1/Mm (880 nm scattering coefficient)