Download Print this page

Sampling Guidelines - Thermo Scientific MIE ADR-1500 Instruction Manual

Particulate monitor

Advertisement

Guidelines and Instrument Layout
Sampling
For ambient air sampling, the omni-directional inlet unit must be used to
minimize wind speed and direction effects on particle sampling
Guidelines
representativeness. This special omni-directional inlet is provided as a
standard accessory of the ADR-1500. On receipt by the customer, this inlet
will arrive packaged separately, and needs to be installed. Refer to
Figure 2–
7
for the final (installed) appearance of the omni-directional inlet. To
install proceed as follows:
Slide the omni-directional inlet onto the inlet stem of the ADR-1500 until
it bottoms. To remove the omni-directional inlet, lift and twist the unit
from the inlet stem.
If the ADR-1500 is to be used for extractive sampling (e.g., from a
chamber, duct, stack, etc.) a flexible plastic tubing (preferably electrically
conductive) can be used and connected to the inlet stem on the upper face
of the ADR-1500. In this case the omni-directional inlet is not used and a
3/8-inch compression fitting with Teflon or nylon ferrules should be used.
For sampling situations involving water sprays, fog, etc. it is recommended
that the in-line inlet heater be switched on. This optional operational
configuration ensures the sample relative humidity will not exceed 70%. If
not necessary, and to extend the run-time of the ADR-1500, when running
off of the internal battery, simply switch the internal heater off.
Equipment Damage Damage may occur to the instrument if the
environmental conditions exceed IP65.
Thermo Fisher Scientific
MIE ADR-1500 Instruction Manual
2-9

Advertisement

loading