Removing The Replenishment Control Valves - Kodak X-Omat 460 RA Processor Service Manual

Hide thumbs Also See for X-Omat 460 RA Processor:
Table of Contents

Advertisement

Removing the Replenishment Control Valves

For this procedure the processor must be deenergized. See page 1-6. In addition, the
RECEIVING-END ACCESS PANEL must be removed from the processor. See pages 1-3
through 1-5 for instructions, if necessary. Also make sure that the water supply to the processor
is shut off for this procedure.
[1]
Place a CLAMP on the INPUT and OUTPUT HOSES of the REPLENISHMENT CONTROL VALVE to be
removed.
[2]
Loosen the HOSE CLAMPS and remove the HOSES from the REPLENISHMENT CONTROL VALVE.
[3]
Remove the HANDLE from the REPLENISHMENT CONTROL VALVE.
[4]
Remove the SCREW securing the BRACKET to the REPLENISHEMENT CONTROL VALVE.
[5]
Remove the REPLENISHMENT CONTROL VALVE from the processor.
[6]
Install the new REPLENISHMENT CONTROL VALVE.
[7]
Install all parts removed in previous steps and make the necessary connection.
FIXER REPLENISHMENT
CONTROL VALVE
5-26
HOSE
CLAMPS
Figure 5-23 Removing the Replenishment Control Valves
IMPORTANT
SCREW
DEVELOPER REPLENISHMENT
CONTROL VALVE
H108_0045BCA
H108_0045BA
636739