Removing the Replenisher Pumps

IMPORTANT
For this procedure the processor must
be deenergized. See page 1-6. In
addition, the DRIVE SIDE ACCESS
PANEL must be removed from the
processor. See pages 1-3 through 1-5
for instructions, if necessary.
[1]
Install CLAMPS onto both of the
REPLENISHMENT TUBES to prevent the
replenishment solution from draining.
[2]
Disconnect either CONNECTOR P/J 37
located at the back of the DEVELOPER
PUMP COVER, or CONNECTOR P/J 38
located at the back of the FIXER PUMP
COVER.
[3]
Remove the 2 MOUNTING SCREWS.
[4]
Remove the REPLENISHER PUMP.
[5]
If necessary, install new parts or, install a new
REPLENISHER PUMP. See Figure 5-4 on
page 5-6 for the location of parts.
[6]
Install all parts removed in previous steps and
make any necessary connections.
636739
DEVELOPER
PUMP
COVER
MOUNTING
CONNECTOR
SCREW (2)
P/J 37
Figure 5-3 Disconnecting Connectors
P/J 37 and P/J 38
H108_0080ACA
H108_0080AA
5-5