Removing the O-Ring from the Recirculation Pump

For this procedure the processor must be deenergized. See page 1-6. In addition, the DRIVE
SIDE ACCESS PANEL must be removed from the processor. See pages 1-3 through 1-5 for
instructions, if necessary.
[1]
Place a CLAMP, TL-2170, on both of the REPLENISHMENT TUBES to prevent leakage of replenishment
solution.
[2]
Remove the 4 WING NUTS and the OUTER PUMP HOUSING. See Figure 5-2 on page 5-4.
[3]
Remove the O-RING SEAL.
[4]
Install a new O-RING SEAL.
[5]
Install all parts removed in previous steps.
[6]
Calibrate the REPLENISHMENT SYSTEM. See the Operator Manual.
[7]
Check the replenishment volumes. See the Operator Manual.
[8]
Remove the CLAMP from each of the REPLENISHMENT TUBES.
636739
IMPORTANT
5-3