[2]
Remove the FILTER CAP.
[3]
Remove the DEVELOPER FILTER.
[4]
Disconnect the 2 HOSES from the FILTER
CANISTER.
[5]
Push the LATCH and lift the FILTER
CANISTER out of the processor.
[6]
Install the new FILTER CANISTER.
[7]
Connect the 2 HOSES to the FILTER
CANISTER.
[8]
Install the new DEVELOPER FILTER and
FILTER CAP.
636739
FILTER
CANISTER
HOSES
Figure 5-8 Removing the Developer
Filter Canister
FILTER
CAP
DEVELOPER
FILTER
LATCH
H108_0116CCA
H108_0116CA
5-11