System Configuration And Open/Short Calibration - Keysight Technologies B1505A User Manual

Power device analyzer / curve tracer
Hide thumbs Also See for B1505A:
Table of Contents

Advertisement

Table 5-2
NOTE
Measurement Examples
I-V and Capacitance Measurements Using N1272A
Connections for the IV Measurement
B1505A
Output Terminal
GNDU
Interlock

System Configuration and Open/Short Calibration

Select selector and measurement resources used for the measurement, and then
measure the Open and Short calibration data of all capacitance measurement paths
including the DUT interface as follows.
1. Click the Configuration button in the EasyEXPERT main window.
The Configuration window appears.
2. Click the Device Capacitance Selector tab to display the Device Capacitance
Selector tab screen.
3. Set up the Device Capacitance Selector and select the used measurement
resources.
To set up an I-V measurement using the IV path of the N1272A Device Capacitance
Selector, select IV in the Default Path field.
4. Click the Start Calibration... button to measure the calibration data of
measurement path. The Device Capacitance Calibration dialog box appears.
If you have measured and saved the calibration data of measurement path, click
the Select Calibration Data... button to select and load the stored calibration
data. Go to the step 7.
5. Click the Start button.
Follow the instruction displayed on the dialog to measure the open and short
calibration data.
It will take about 20 minutes to obtain the calibration data.
6. After finishing the calibration, click the Save and Apply button to save and apply
the calibration data.
7. Click the Close button to close the dialog box.
5-26
Cable
GNDU Cable
GNDU
Interlock Cable
Interlock
Keysight B1505A User's Guide, Edition 12
N1265A
Input Terminal