CAUTION
Figure 3-12
Accessories
•
Selector Output Low to DUT low (ex. Emitter/Source)
•
Gate High to DUT drive (ex. Base/Gate Force)
•
Gate Low to DUT low (ex. Emitter/Source Force)
HC/MC/DHCSMU via SMU 3 terminals should be connected as follows.
•
SMU 3 High Force and Sense to DUT high
•
SMU 3 Low Force and Sense to DUT low
•
GNDU Force and Sense to DUT low
The GNDU signals will appear at the Selector Output Low terminals as
shown in
Table
3. Set the DUT on the socket.
4. Close the fixture cover and perform measurement.
Do not apply voltage/current over the maximum limit of the socket module.
Inline Package Socket Module
165 mm
N1265A Opt 010
Emi er / Source
Collector / Drain
2
F
S
F
2
1
2
1
5
3
4
Inline Package Socket (3 Pin)
14 mm
3.48 mm
1.78 mm
3-46
3-5.
3
1
Base / Gate
S
F
S
3
165 mm
6
Height: 34 mm
(45 mm including terminal)
(58 mm including socket)
Keysight B1505A User's Guide, Edition 12
Maximum voltage: 3000 V
Maximum current:
1-F: 500 A pulse, 39 A dc
1-S: 40 A pulse, 2 A dc
2-F: 500 A pulse, 39 A dc
2-S: 40 A pulse, 2 A dc
3-F: 500 A pulse, 39 A dc
3-S: 40 A pulse, 2 A dc
F: Force
S: Sense