Table 2-17
Range
Output Value
1500 V
0 |V| 1500 V
3000 V
0 |V| 2200 V
a. Pulse base value is always 0 V.
b. Measurement is performed by the 3000 V measurement range.
Table 2-18
Range
Output Value
100 mA
2 A
Introduction
Measurement Resources
HVMCU Voltage Ranges and Resolutions
Measurement
a
Value
0 |V| 2200 V
HVMCU Current Ranges and Resolutions
Measurement
Value
0 |I| 0.11 A
0 |I| 1.1 A
0 |I| 2.5 A
2-42
Setting/
Measurement
b
Resolution
1.5 mV/3 mV
3 mV/3 mV
Setting
Resolution
Keysight B1505A User's Guide, Edition 12
Compliance
Maximum
Range
Current
2 A
b
1 A
100 mA
± 110 mA
Measurement
Maximum
Resolution
200 nA
± 2200 V
4 A
± 2200 V
4 A
± 1500 V
± 2.5 A
± 1.1 A
Voltage