Table 3-6
Connector
label
High Force
Low Force
High Sense
Low Sense
Gate
a. Pulse of 1 ms pulse width and 0.4 % duty cycle
WARNING
CAUTION
Accessories
Prober System Cable (N1254A-524)
This is the cable for extending the N1265A's Selector Output and Gate terminals
to a prober station. Cable length is 1.8 m. Prober side connector type and the
maximum voltage/current are shown in
This cable internally connects Selector Output Low Sense and Gate Low.
Prober Side Connectors and Maximum Voltage/Current of N1254A-524
Connector
N1265A output terminals
type
Banana(jack)
Selector Output High Force
Banana(jack)
Selector Output Low Force
HV(jack)
Selector Output High Sense
BNC(f)
Selector Output Low Sense
Gate Low
BNC(f)
Gate High
For using the prober system cable safely
Never connect the interlock cable to the N1265A.
Never leave and use the N1265A in the condition that the clear plastic plates
are removed from its fixture cover.
Pour utiliser le système de sonde en toute sécurité
Ne jamais connecter le câble de verrouillage au N1265A.
Ne jamais ôter les plaques en plastique transparent du couvercle d'appareil du
N1265A.
Do not apply voltage/current over the maximum limit of the cable.
3-54
Table
3-6.
Maximum
to extend
Keysight B1505A User's Guide, Edition 12
Maximum
voltage
current
39 A dc,
3000 V
500 A pulse
200 V
1 A dc,
3000 V
20 A pulse
200 V
200 V
a