CAUTION
Figure 3-3
If MCSMU, HCSMU, or DHCSMU is used, High Force and Sense must be
connected to Force and Sense used for the high terminal of a device under
test (DUT), respectively. Low Force and GNDU Force must be connected to
Force used for the low terminal of the DUT, and Low Sense and GNDU
Sense must be connected to Sense used for the low terminal of the DUT.
c. Set the DUT on the socket.
d. Close the fixture cover and perform measurement.
Do not apply voltage/current over the maximum limit of the socket module.
Inline Package Socket Module
165 mm
Force
1
Sense
14 mm
3.48 mm
1.78 mm
5. Universal Socket Module (N1259A-011)
This is a blank module, kind of a do-it-yourself kit for supporting variety of
packaged devices. This module can be used by mounting your desired socket or
packaged device and making connections same as the N1259A-010 Inline
Package Socket Module.
To use this module, see the following simple instruction. For the component
locations and dimensions, see
Module (N1259A-010)" on page 3-18
Keysight B1505A User's Guide, Edition 12
Height: 34 mm
Force
(45 mm including terminal)
2
(58 mm including socket)
Sense
165 mm
Force
Maximum voltage: 3000 V
3
Sense
Maximum current: 40 A pulse, 2 A dc
Figure
3-4. Also see
"Inline Package Socket
to perform measurement.
Accessories
3-19