Figure 3-33
CAUTION
Accessories
Curve Tracer Test Adapter Socket Module (N1273A-013)
This module provides a socket available for connecting a test adapter designed
for connecting to Tektronix 370B/371B curve tracers. The socket module
internal connection is shown in
Curve Tracer Test Adapter Socket Module
Emitter / Source
Collector / Drain
AC/DC Guard
2
1
5
3
4
To use this module, see the following simple instruction.
1. Connect the socket module to the Base/Gate, Collector/Drain, and
Emitter/Source connectors on the test fixture directly.
2. Connect your test adapter to the socket.
3. Set the DUT on your test adapter.
4. Close the fixture cover and perform measurement.
Do not apply voltage/current over the maximum limit of the socket module.
Terminal related voltage: 3000 Vdc
Terminal related current: 100 mA
3-96
Figure
3-33.
Base / Gate
6
Keysight B1505A User's Guide, Edition 12
Internal connections
Socket Terminal
N1273A Output
1: Collector/Drain Force
Collector/Drain
2: Collector/Drain Sense
No Connection
3: Emitter/Source Force
Emitter/Source
4: Emitter/Source Sense
No Connection
5: Base/Gate Force
Base/Gate
6: Base/Gate Sense
No Connection