Using Multiple Hvsmu Modules; Hvsmu Output Range - Keysight Technologies B1505A User Manual

Power device analyzer / curve tracer
Hide thumbs Also See for B1505A:
Table of Contents

Advertisement

Using Multiple HVSMU Modules

This section describes how to perform source output and measurement using the
B1505A installed with multiple HVSMU modules.

"HVSMU Output Range"

"Setting Output Range Mode"
"Performing Measurement"
"Available Application Test List"
HVSMU Output Range
If the B1505A is installed with multiple HVSMU modules, specify the output range
mode first to enable the HVSMU. The following lists the available modes.
Single channel mode (initial setting)
An HVSMU channel is available. The other HVSMUs work as the channels of 0
V output with 4 mA compliance.
±1500 V mode
The output range of all HVSMUs is from -1500 V to 0 V and from -8 mA to 0
mA, or from 0 V to +1500 V and from 0 mA to +8 mA.
+3000 V mode
The output range of all HVSMUs is from 0 V to +3000 V and from 0 mA to +4
mA.
-3000 V mode
The output range of all HVSMUs is from -3000 V to 0 V and from -4 mA to 0
mA.
Keysight B1505A User's Guide, Edition 12
Measurement Examples
Using Multiple HVSMU Modules
5-45