CAUTION
Figure 3-25
NOTE
Output Low
SHV(jack) connector. Use SHV(plug)-SHV(plug) cable of 16493V to connect
N1265A test fixture.
To connect prober, use N1269A adapter and SHV(plug)-SHV(plug) cable of
16493V. See
"N1269A Ultra High Voltage Connection Adapter" on page 3-70
for connecting a prober.
Output High
UHV(jack) connector. Use UHV(plug)-UHV(plug) cable of 16493V to connect
N1265A test fixture.
To connect prober, use ultra high voltage cable with UHV(plug) connector,
manipulator, and such.
Never connect the UHVU High terminal to any output, including circuit common,
chassis ground, or any other measurement resource such as SMU. Connecting other
measurement resource may damage the connected one.
To Configure UHVU
B1505A
MCSMU
MC/HCSMU
Digital I/O
Interlock
For the gate terminal of DUT, use MCSMU. Another module cannot be used.
Keysight B1505A User's Guide, Edition 12
N1268A UHV expander
I-control
V-control
High
Digital I/O
Low
Interlock Input
Interlock Output
Accessories
DUT
Drain
Source
Interlock circuit
3-69