Table 2-7
Output Value and
Range
Measurement Value
0.2 V
0  |V|  0.2 V
2 V
0  |V|  2 V
20 V
0  |V|  20 V
40 V
0  |V|  40 V
Table 2-8
Output Value and
Range
Measurement Value
10 A
0  |I|  11.5 A
100 A
0  |I|  115 A
1 mA
0  |I|  1.15 mA
10 mA
0  |I|  11.5 mA
100 mA
0  |I|  115 mA
1 A
0  |I|  1.15 A
2 A
0  |I|  2.02 A
20 A
0  |I|  20 A
40 A
0  |I|  40 A
a. Maximum value is 1 A for DC output or measurement by HCSMU.
b. Maximum value is 2 A for DC output or measurement.
NOTE
HCSMU Voltage Ranges and Resolutions (HCSMU / DHCSMU)
Setting/Measurement
HCSMU Current Ranges and Resolutions
Setting/Measurement
a
b
40 A range and 2 A range
The 40 A range and the 2 A range are available when two HCSMU modules are
installed in the B1505A and connected to the 16493S-020 Dual HCSMU Kelvin
combination adapter or the 16493S-021 Dual HCSMU combination adapter. Two
HCSMU modules can be the DHCSMU which supports ± 40 A (pulse), ± 2 A (DC).
Then the 20 A range is not supported.
When the compliance value is set to > 1 A or < -1 A, the current measurement
range is fixed to the compliance range.
Keysight B1505A User's Guide, Edition 12
Resolution
200 nV
± 20 A / ± 40 A
2 V
± 20 A / ± 40 A
20 V
± 20 A / ± 40 A
40 V
Resolution
10 pA
100 pA
1 nA
10 nA
100 nA
1 A
2 A
20 A
40 A
Measurement Resources
Maximum
Maximum Pulse
Current
Base Value
± 0.2 V
± 20 V
± 1 A / ± 2 A
± 40 V
Maximum
Maximum Pulse
Voltage
Base Value
± 40 V
± 10 A
± 40 V
± 100 A
± 40 V
± 1 mA
± 40 V
± 10 mA
± 40 V
± 100 mA
± 40 V
± 40 V
± 20 V
± 100 mA
± 20 V
± 200 mA
Introduction
± 2 V
± 1 A
± 2 A
2-33