CAUTION
7. Connect one end of all cables to the inside terminals on the blank module.
Remove screws and tooth washers from the terminals, and fix the lug
terminal of the cable end in order of a washer, a lug terminal, a tooth washer,
and a screw. When fixing the lug terminal, secure a screw by a T15 Torx
screwdriver with fixing the bottom nut by a 9 mm wrench.
The following shows the positions of the inside terminals in the blank
module.
Blank Module
(Wiring side view)
8. Connect the other end of the cables to the terminals on the back side of the
socket board.
If a socket terminal does not have a screw or nut for securing a lug terminal
of the cable, prepare it. Ensure the terminal position of the blank module and
the socket board, and connect the cables correctly.
9. Fix the bottom cover to the blank module using a T10 Torx screwdriver.
Do not apply voltage/current over the maximum limit of the socket module.
Terminal rated voltage: 3000 Vdc
Terminal rated current: 120 mA