Keysight Technologies B1505A User Manual page 188

Power device analyzer / curve tracer
Hide thumbs Also See for B1505A:
Table of Contents

Advertisement

Figure 3-30
CAUTION
Accessories
Une tension dangereuse, une tension de sortie maximale de l'appareil peut
apparaître aux bornes de mesure si le couvercle de l'équipement est fermé.
3-pin Inline Package Socket Module (supplied with the N1273A)
Use the socket module to connect a three-terminal inline packaged device.
3-pin Inline Package Socket Module
DUT
Test fixture
A short plate is furnished with the socket module. It is used for performing the
short correction before the capacitance measurement. Set it on the socket before
the short correction and remove it after the correction.
To use this module, see the following simple instruction.
1. If the earthing wire is not connected to the socket module, connect it.
2. Attach the socket module directly to the Base/Gate, Collector/Drain, and
Emitter/Source terminals of the test fixture.
3. Connect the earthing wire of the socket module to the chassis common
terminal.
4. Set your DUT on the socket.
5. Close the fixture cover and perform measurement.
Do not apply voltage/current over the maximum limit of the socket module.
Terminal rated voltage: 3000 Vdc
Terminal rated current: 120 mA