Keysight Technologies B1505A User Manual page 252

Power device analyzer / curve tracer
Hide thumbs Also See for B1505A:
Table of Contents

Advertisement

WARNING
Measurement Examples
Temperature Controlled by Thermal Plate
To prevent electrical shock and DUT damage, do not connect or disconnect the
DUT while the instrument is applying voltage or current.
When you touch the DUT after measurement, devise a countermeasure of
residual charge and heat to prevent electrical shock and burn. Use gloves and
any tool. Also have enough time for discharge and radiation.
Afin d'éviter toute décharge électrique et dommage MST, ne branchez ou
débranchez pas la sortie MST alors que l'appareil envoie de la tension ou du
courant.
Lorsque vous touchez le MST après la mesure, élaborez une contre-mesure de
la charge résiduelle et du chauffage afin d'éviter tout choc électrique et toute
brûlure. Utilisez des gants et des outils. Prévoyez également du temps pour la
décharge et la radiation.
Prepare a measurement as follows:
1. Connect the test fixture, the socket module, and the required accessories.
For more information, see the following sections:
"N1259A Test Fixture" on page 3-9
"N1265A Ultra High Current Expander/Fixture" on page 3-33
"N1271A-001 Thermal Plate Compatible Enclosure for N1259A/N1265A"
on page 3-71
2. Set up the thermal plate and the N1271A-001 enclosure on the N1259A or
N1265A test fixture.
3. Connect a GPIB cable between the controller (B1505A or external PC) and the
thermal plate.
4. Prepare a DUT (device under test).
5. Turn the instrument on.
Do not connect the device at this time.
6. Perform the GPIB settings for controlling the thermal plate.
Set the GPIB interface on the controller (B1505A or external PC) as
"System Controller".
Check the GPIB addresses on the controller and the thermal plate.
If they are duplicated, set unique addresses.
5-34
Keysight B1505A User's Guide, Edition 12