Figure 3-13
N1265A Opt 011
Collector / Drain
Force
82 mm
(Nut pitch: 69 mm)
3
4
91 mm (Nut pitch: 78 mm)
Universal Socket Module (N1265A-011)
This is a blank module, kind of a do-it-yourself kit for supporting variety of
packaged devices. This module can be used by mounting your desired socket or
packaged device and making connections same as the N1265A-010 Inline
Package Socket Module.
To use this module, see the following simple instruction. For the component
locations and dimensions, see
Module (N1265A-010)" on page 3-45
Universal Socket Module
Emi er / Source
Base / Gate
Sense
Force
Sense
Force
Sense
1
2
2
1
5
6
7
3
4
8
Universal Socket Module
1. Prepare the following parts.
blank board suitable for mounting the socket or packaged device
screw M3, 4 ea., for fixing the blank board on the socket module
wire, adequate length and quantity, for making connections
hexlobe (torx type) screwdriver T-10
socket, if you use, and packaged device under test (DUT)
2. Cut the blank board in 90 mm  81 mm square.
3. Make four screw holes on the board. The holes should be 6 mm inside from
the edge.
4. Fix the board to the blank module.
5. Remove the cover bottom of the blank module.
6. Mount the socket or DUT on the board and solder wire between its terminals
and the blank module terminals.
Keysight B1505A User's Guide, Edition 12
Figure
3-13. Also see
to perform measurement.
5
6
129 mm
"Inline Package Socket
165 mm
160 mm
Accessories
165 mm
3-47

Hide quick links: