WARNING
CAUTION
Accessories
Make enough space between the High-Low adapters, also between the high side
adapter and the shield/chassis, for example, about 1 mm for maximum 200 V
output and 6 mm for 3000 V, to prevent discharge and any accident.
Laissez suffisamment d'espace entre les adaptateurs High-Low et également
entre l'adaptateur du côté élevé et la protection/le châssis. Par exemple, environ
1 mm pour une sortie de 200 V au maximum et 6 mm pour 3 000 V afin d'éviter
toute décharge et tout accident.
f. Close the fixture cover and perform measurement.
7. Curve Tracer Test Adapter Socket Module (N1259A-013)
This module provides a socket available for connecting a test adapter designed
for connecting to Tektronix 370B/371B curve tracers. Socket module internal
connection is shown in
To use this module, see the following simple instruction.
a. Attach the socket module to the test fixture.
b. Connect your test adapter to the socket.
c. Connect wires (N1254A-508 or 509) between the socket module terminals
and the fixture output terminals.
For making the Kelvin connection, Force and Sense must be connected to
Force and Sense of the socket module respectively.
If MCSMU, HCSMU, or DHCSMU is used, High Force and Sense must be
connected to Force and Sense used for the high terminal of a device under
test (DUT), respectively. Low Force and GNDU Force must be connected to
Force used for the low terminal of the DUT, and Low Sense and GNDU
Sense must be connected to Sense used for the low terminal of the DUT.
d. Set the DUT on your test adapter.
e. Close the fixture cover and perform measurement.
Do not apply voltage/current over the maximum limit of the socket module.