Keysight Technologies B1505A User Manual page 141

Power device analyzer / curve tracer
Hide thumbs Also See for B1505A:
Table of Contents

Advertisement

WARNING
To prevent electrical shock and DUT damage, do not connect or disconnect the
DUT while the instrument is applying voltage or current.
When you touch the DUT after measurement, devise a countermeasure of
residual charge and heat to prevent electrical shock and burn. Use gloves and
any tool. Also have enough time for discharge and radiation.
Afin d'éviter toute décharge électrique et dommage MST, ne branchez ou
débranchez pas la sortie MST alors que l'appareil envoie de la tension ou du
courant.
Lorsque vous touchez le MST après la mesure, élaborez une contre-mesure de
la charge résiduelle et du chauffage afin d'éviter tout choc électrique et toute
brûlure. Utilisez des gants et des outils. Prévoyez également du temps pour la
décharge et la radiation.
Inline Package Socket Module (N1265A-010)
This module provides a socket used for connecting three-terminal inline
packaged device and three couples of the Force and Sense terminals. Socket
module internal connection is shown in
Short bar is furnished with the module. It is used for performing the short
correction before the impedance measurement. Set the short bar before the short
correction and remove it after the correction.
To use this module, see the following simple instruction.
1. Attach the socket module to the test fixture.
2. Connect wires between the socket module terminals and the fixture output
terminals. Then use the following wire (furnished with the N1265A-010).
180 mm length wire (yellow) for Selector Output High/Low Force
180 mm length wire for Selector Output High/Low Sense
250 mm length wire for Gate, SMU, or chassis
For making the Kelvin connection, Force and Sense must be connected to
Force and Sense of the socket module respectively.
For the high voltage capacitance measurement, use SHV(plug)-SHV(plug)
cable (N1254A-512) and SHV(jack)-banana adapter (N1254A-513) for
connection.
Selector Output and Gate should be connected as follows.
Selector Output High to DUT high (ex. Collector/Drain)
Keysight B1505A User's Guide, Edition 12
Figure
3-12.
Accessories
3-45

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents