WARNING
WARNING
WARNING
Accessories
Warming labels written in French, German, and Japanese are furnished. Attach
the label to the front panel of the fixture if you need.
The red light indicates that hazardous voltage (maximum ± 3000 Vdc) may
appear at measurement terminals. Check this indicator before accessing.
Le témoin rouge indique qu'une tension dangereuse (maximum ± 3000 V CC)
risque d'apparaître au niveau des bornes de mesure. Vérifiez cet indicateur
avant d'accéder.
4. Fixture cover
The fixture cover should be closed to avoid electrical shock by touching
measurement terminals and to prevent a device under test from external noise.
When the fixture cover is open, maximum SMU output is limited to ± 42 V.
Hazardous voltage, instrument maximum output voltage may appear at the
Force, Guard, and Sense terminals if the fixture cover is closed.
Une tension dangereuse, une tension de sortie maximale de l'appareil peut
apparaître aux bornes Force, Guard et Sense si le couvercle de l'équipement est
fermé.
Make sure that the cover is closed properly before starting measurement. Do
not perform the measurement when a wire is protruding from the fixture cover.
Assurez-vous que le couvercle est fermé correctement avant de commencer la
mesure. Ne pas effectuer la mesure lorsqu'un câble dépasse du couvercle de
l'appareil.
3-10
Keysight B1505A User's Guide, Edition 12