Programmable DC Electronic Load 63600 Series Operation & Programming Manual
to none, master or slave for parallel run.
NONE:
Disable the MASTER/SLAVE control function.
MASTER:
Used as the master for the parallel group and this is the only one controlled by
front panel or PC in this group. Also tell the slaves how many current they
should sink. Slave Model: 1 – 5 to setup the slave's model to use in parallel.
NONE means not exist.
SLAVE:
Setup the load as slave.
To leave out of the Synchronous Dynamic mode edit page of system configuration, you need
ADVA
to press
and
1.
2.

4.7.6 Recall Factory Default

In the page of system configuration, turn the Rotary knob to change the display value to 7,
ENTER
then press
key into Recall Factory Default page of system configuration.
Set Recall the load Factory default When you choose YES and press
recall the factory default setting as Table 4-5 shows.
To leave out of the Recall Factory Default page of system configuration, you need to press
ENTER
ADVA
and
Mode Of
Range
Operation
H
CC
M
L
H
CR
M
L
4-44
ENTER
simultaneously to go back to the page of system configuration.
In the page of system configuration, turn the Rotary knob to change
the display value to 6, then press
of system configuration. Normally, we recommend that normal
users don't enter into this page and edit the data. This is for
Chroma instruments factory or service center or standard
instruments calibration center to calibrate the programming and
measurement values that are out of accuracy of specifications.
The fastest refresh time for LOAD panel is 0.5 second. When
operating in SYNC DYNA, if the T1 or T2 time is less than 0.5
second, the change of panel reading is restricted by the panel
refresh time. Thus it may not seem to be synchronized but it dose in
actual loading.
[ DEFAULT ]
RECALL FACTORY
DEFAULT : YES
simultaneously to go back to the page of system configuration.
Table 4-5 Factory Default
63640-80-80
63630-80-60
00.005A
00.004A
0.0005A
0.0004A
000.05mA
000.04mA
2900.0Ω
3000.0Ω
720.00Ω
600.00Ω
020.00Ω
30.000Ω
ENTER
key into Calibration page
63630-600-
63610-80-20
15
00.001A
0.0000A
0.0001A
000.00mA
000.01mA
00.000mA
12000Ω
200.00kΩ
2900.0Ω
4000.0Ω
080.00Ω
270.00Ω
ENTER
, the Load will
63640-150-
60
00.000A
0.0000A
000.00mA
1500.0Ω
800.00Ω
60.000Ω