Programmable DC Electronic Load 63600 Series Operation & Programming Manual
2.
Set Load Level. There are 15,000 discrete steps from 0 to full scale in each range and
each mode. Turn the Rotary knob to change the display value to 10.000A, then press
ENTER
key to confirm.
3.
Set Slew Rate. Setting the rising and falling slew rate. Turn the Rotary knob to change
the display value, then press
4.
Set Trigger Mode. There are three Trigger Modes and they are RISE , FALL and
HOLD_UP. Turn the Rotary knob to change the mode until the desired mode is
displayed on the VFD. Then, press
5.
Set Trigger Voltage. Trigger Voltage is the conduction voltage level.
TRG_S: It sets the start trigger voltage level for measurement time.
TRG_E: It sets the end trigger voltage level for measurement time.
The Electronic Load will measure the duration from the load on to the UUT output
voltage equal to the setting trigger voltage, and the Load stops sinking current when the
UUT output down to reach the voltage. Turn the Rotary knob to change the display
value, then press
6.
Set the period of time out. The Electronic Load will measure the duration from the load
on to the UUT output voltage equal to the setting trigger voltage. When the time is
already over the period of time out, but the UUT output voltage still isn't achieve to the
trigger voltage, the Load will load off and stop counting the timing. Turn the Rotary knob
to change the display value, then press
4-24
MODE: CC
I_SET: 10.000 A
ENTER
key to confirm.
: 0.0012A/μs
: 0.0012A/μs
ENTER
TRG_M: FALL
ENTER
key to confirm.
TRG_S :3.000
TRG_E :5.000
T_OUT: 600 s
key to select mode and confirm the setting.
V
V
ENTER
key to confirm.