Checking Dynamic Constant Current High Range; Checking Dynamic Constant Current Medium Range - Chroma 63610-80-20 Operation & Programming Manual

Programmable dc electronic load
Table of Contents

Advertisement

7.3.5.1 Checking Dynamic Constant Current High Range

MODE
A.
First, press
indicator.
EDIT
B.
Press
to enter into Dynamic Constant Current setting. The settings are listed in
Table 7-17.
C.
Set the DC Source output voltage to 5V (hint: Set to 7V for Model 63630-600-15), the
limit current needs to be larger than the highest level set by dynamic constant current.
(EX: When the 63630-600-15 H level is set to 15A, the DC Source limit current can set
to 17A.)
When the 63640-150-60 is doing dynamic loading, the minimum working
voltage should be larger than 2.5V.
Model
63630-600-15
63640-150-60
63630-80-60
63610-80-20
63640-80-80

7.3.5.2 Checking Dynamic Constant Current Medium Range

MODE
A.
First, press
indicator.
EDIT
B.
Press
to enter into Dynamic Constant Current setting. The settings are listed in
Table 7-18.
C.
Set the DC Source output voltage to 5V and the limit current needs to be larger than the
highest level set by dynamic constant current. (EX: When the 63630-600-15 H level is
set to 1.5A, the DC Source limit current can set to 2A.)
Model
63630-600-15
63640-150-60
63630-80-60
63610-80-20
63640-80-80
till VFD shows CC and press
Table 7-17
H
L
15A
0A
60A
1A
60A
0A
20A
0A
80A
0A
till VFD shows CC and press
Table 7-18
H
L
1.5A
0A
6A
0A
6A
0A
2A
0A
8A
0A
RANGE
to light up the H range LED
T1
T2
10ms
10ms
10ms
10ms
10ms
10ms
10ms
10ms
10ms
10ms
RANGE
to light up the M range LED
T1
T2
10ms
10ms
10ms
10ms
10ms
10ms
10ms
10ms
10ms
10ms
Verification
SR/
SR\
1.5 A/μs
1.5 A/μs
6 A/μs
6 A/μs
6 A/μs
6 A/μs
2 A/μs
2 A/μs
8 A/μs
8 A/μs
SR/
SR\
0.15 A/μs 0.15 A/μs
0.6 A/μs
0.6 A/μs
0.6 A/μs
0.6 A/μs
0.2 A/μs
0.2 A/μs
0.8 A/μs
0.8 A/μs
7-13