Page 1
Modular DC Power Supply 62000B Series User’s Manual...
Page 3
Modular DC Power Supply 62000B Series Operation Manual Version 1.5 July 2013...
Page 4
The information in this document is subject to change without notice. Chroma ATE INC. makes no warranty of any kind with regard to this manual, including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose.
Page 5
Chroma’s obligation under this warranty is limited solely to repairing any such instrument, which in Chroma’s sole opinion proves to be defective within the scope of the warranty when returned to the factory or to an authorized service center. Purchaser is responsible for the shipping and cost of the service item to Chroma factory or service center.
Page 6
Material Contents Declaration The recycling label shown on the product indicates the Hazardous Substances contained in the product as the table listed below. : See <Table 1>. : See <Table 2>. <Table 1> Hazardous Substances Lead Mercury Cadmium Hexavalent Polybrominated Polybromodiphenyl Part Name Chromium...
Page 7
“ ” indicates that the level of the specified chemical substance exceeds the threshold level specified in the standards of SJ/T-11363-2006 and EU 2005/618/EC. Chroma is not fully transitioned to lead-free solder assembly at this moment; however, most of the components used are RoHS compliant.
Page 9
Failure to comply with these precautions or specific WARNINGS given elsewhere in this manual will violate safety standards of design, manufacture, and intended use of the instrument. Chroma assumes no liability for the customer’s failure to comply with these requirements.
Page 10
Safety Symbols DANGER – High voltage. Explanation: To avoid injury, death of personnel, or damage to the instrument, the operator must refer to an explanation in the instruction manual. High temperature: This symbol indicates the temperature is now higher than the acceptable range of human. Do not touch it to avoid any personal injury.
Page 11
− “Installation in Rack” in the chapter of “Installation”. − “Setting Over Voltage Protection” and “Application & Control Method for 62000B Series DC Power Supply” in the chapter of “System Operation & Usage”. − “Rule for Setting 62000B CAN BUS Parameter (Slave)”, “Rule for Setting CAN Adapter Parameter (Master)”, “Description of CAN...
Remote Sensing ....................3-4 3.5.1 Correct Usage....................3-4 3.5.2 When Connecting Remote Sensing Wire Polarity Reversely ......3-5 Mainframe Parallel Connection................3-5 Application & Control Method for 62000B Series DC Power Supply .....3-7 Communication Protocols .................4-1 Preface ........................4-1 Introduction to Editing....................4-1 4.2.1 Conventions....................4-1 4.2.2 Numerical Data Format..................4-1...
Page 14
Modular DC Power Supply 62000B Series Operation Manual CAN BUS Configuration ..................4-7 4.4.1 Limitations of Communication Protocol ............4-7 4.4.2 Rule for Setting 62000B CAN BUS Parameter (Slave)........4-7 4.4.3 Rule for Setting CAN Adapter Parameter (Master)........4-8 4.4.4 Description of CAN Packet ................4-8 4.4.5...
Graphic Soft Panel Control and Monitor Remote ON/OFF Signal Remote CAN Bus Interface Control DC OK Signal Output The specifications of 62000B Series DC Power Supplies are shown in the table below. (Test condition: 25 ± 5°C and under resistance load.)
Page 16
Modular DC Power Supply 62000B Series Operation Manual SPECIFICATIONS Model 62015B-15-90 62015B-30-50 62015B-60-25 62015B-80-18 62015B-150-10 Electrical Specifications Output Ratings Output power 1350W 1500W 1500W 1440W 1500W Output Voltage 1~15V 1~30V 1~60V 1~80V 1~150V Output Current 1~90A 1~50A 1~25A 1~18A 1~10A Line Regulation 0.1% FS...
Page 17
Overview 25%. Six Position Mainframe through CAN. OVP Protection, please refer to section 3.3.1 Setting Over Voltage Protection. Typical efficiency at 250VAC input voltage and full output power.
Overview 1.4 Function Keys 1.4.1 Front Panel Figure 1-1 Front Panel of 62000B-6-1 Figure 1-2 Front Panel of 62000B-3-1 Item Name Description Voltage V_SET It adjusts the output to set voltage. (The setting is completed after the LED blinked 3 times during adjustment.) Current I_SET It adjusts the output to set current limit.
Modular DC Power Supply 62000B Series Operation Manual 1.4.2 Rear Panel Figure 1-3 Rear Panel of 62000B-6-1 Figure 1-4 Rear Panel of 62000B-3-1...
Page 23
Overview CANBUS TERMINAL R Dummy Terminal CAN_L CAN_H Resistor Local or Remote Setting ON/OFF From left to right CC /CV Setting Pin2 CAN_H Pin5 CAN_L Figure 1-5 Communication Contacts on the Rear Panel Item Name Description Current Sharing It is the connection of Current Sharing in between Main Frames.
Installation 2.1 Checking the Package (1) Check if there is any damage or any missing accessories after unpacking it. (2) Should any damage it found, contact “Chroma RMA” immediately to request return shipment. The machine package is shown as below.
Modular DC Power Supply 62000B Series Operation Manual 2.2 Preparation for Use (1) Be sure the Power Supply is connected to the AC line input that meets the specification. (2) The instrument must be installed in an air-circulated area to avoid the internal temperature getting too high.
Installation 2.3 Installation in Rack 2.3.1 62000B-6-1 For 19” rack each 4U can place one 62000B-6-1 PS: It is suggested to use the support angle of 1.6 mm (SPCC/SECC) for rack mount of 4UH. SECC: Steel - Electrogalvanized - ColdRolled - Coil SPCC: Steel Plate Cold rolled Commercial...
As to output, it is determined by front panel switch. To cope with the various requirements from customers, the 62000B Series Power Supply has 3 kinds of on/off operating modes based on the interface used. All of them can operate the system via 1.
Modular DC Power Supply 62000B Series Operation Manual 3.2 Notices When Connecting Output (1) It is necessary to ensure the withstand current of output cable is larger than the maximum current of LOAD. (2) To prevent the output cable from damage or over-shoot generated by cable inductance,...
System Operation & Usage Install additional fuse 62000B DC Power System The recharge energy is huge that may cause injury to person or damage WARNING to hardware device when the battery is short or other abnormal condition is occurred. It is necessary to install additional fuse to protect the safety of personnel and equipment when applying in battery charge/discharge.
Modular DC Power Supply 62000B Series Operation Manual Power Off Procedure: Turn the MAINFRAME ON/OFF switch to OFF. Set the AC Input to OFF. 3.3.1 Setting Over Voltage Protection Module Description 1 ~ MAX SET VALUE 115% OF SET VALUE...
System Operation & Usage Though the Remote sensing function is able to compensate 3V voltage drop, the maximum output power of this series DC power supply is the rated power (which is the sum of voltage multiplies the current on the output terminal), the power that goes beyond the DC Module can supply will not be compensated with the line voltage drop up to 3V.
Page 36
Modular DC Power Supply 62000B Series Operation Manual Current Sharing CANBUS Wire Wire V Sense Wire Figure 3-3 The Remote Sensing Wire won’t be damaged due to wrong polarity CAUTION connection at this time. Just follow the steps below: First shut down the device.
Fault LEDs are always on from the beginning it means the power module is not inserted properly. Please take out the power module and reinsert it after the Fault LED is off. 3.7 Application & Control Method for 62000B Series DC Power Supply...
Page 38
Modular DC Power Supply 62000B Series Operation Manual...
Communication Protocols Communication Protocols 4.1 Preface The 62000B Series is the DC Power Supply that suitable for burn in test. The communication interface of physical layer is CAN BUS with the Standard Commands for Programmable Instruments (SCPI) applied. 4.2 Introduction to Editing All commands and response messages should be transmitted in ASCII code format.
Modular DC Power Supply 62000B Series Operation Manual 4.2.3 Basic Definition 4.2.3.1 Command Tree Structure The commands of the DC Power Supply are based on a hierarchical structure, also known as a tree system. In order to obtain a particular command, the full path to that command must be specified.
Communication Protocols 4.2.3.8 Program Message Unit Separator (;) The separator (semicolon ;) separates the program message unit elements from one another in a program message. Example: VOLT 80; CURR 15<PMT> 4.2.3.9 Program Message Terminator (<PMT>) A program message terminator represents the end of a program message. The new command line is a single byte 0Ah (line feed) in ASCII code.
Modular DC Power Supply 62000B Series Operation Manual 4.3 Commands of DC Power Supply This section describes the syntax and parameters of all commands for DC Power Supply. (1) *CLS Clear Status Type: Device status Description: *CLS command acts the follows: Clear Error Code and Reset Error Message.
Page 43
Communication Protocols (6) CONFigure: BAUDrate Description: It sets the baudrate (default is 125000). Syntax: CONFigure: BAUDrate <NR1>[suffix] Parameter: 10000, 20000, 50000, 100000, 125000, 250000, 500000, 800000, 1000000 Example: CONFigure:BAUD 250000 It sets the baudrate to 250kbps. Once the 62000B DC Power Supply received this command, the baudrate will Note change immediately to the new setting.
Page 44
Modular DC Power Supply 62000B Series Operation Manual (12) FETCh: CURRent? Description: It measures the output of power supply module and returns the measured current. Query Syntax: FETCh:CURRent? Return Parameter: <NR2> [Unit=Amp] Query Example: FETC:CURR? Return Example: 3.15 (13) FETCh: STATus? Description: It returns the status code of power supply’s state.
The Master address cannot be the same as 62000B Power Supply module. Server/Client Structure: The power module of 62000B Series is Slave and it does not send message actively but use the way of question and answer. Broadcasting Message: The Slave executes the broadcasted message without responding it.
Page 48
Modular DC Power Supply 62000B Series Operation Manual Destination Address: Source Address: 254 RTR=0 DLC=8 DATA=8 Bytes (PC) (Data Packet) (Data Length) 0 0 0 0 0 0 0 1 1 1 1 1 1 1 1 0 x x x x x x x x x x x x x ‘S’...
Communication Protocols Example: FETC:CURR? Fetch mainframe output current (Address 1) ID=0000000100100000xxxxxxxxxxxxxb DCL=8 RTR=0 DATA=" FETC:CUR " +0x0A Destination Address: Source Address: 254 RTR=0 DLC=8 DATA=8 Bytes (PC) (Data Packet) (Data Length) 0 0 0 0 0 0 0 1 1 1 1 1 1 1 1 0 x x x x x x x x x x x x x ‘F’...
Page 50
Modular DC Power Supply 62000B Series Operation Manual ‘CAN Port Status Define Const RxBuffer = 1 Const DataOverrun = 2 Const TxBuffer = 4 Const TxEnd = 8 Const RxStatus = 16 Const TxStatus = 32 Const ErrorStatus = 64...
Page 54
Modular DC Power Supply 62000B Series Operation Manual TxBuf(9)= &H3F ‘? TxBuf(10)= &HA ‘line feed Can_Tx TxBuf,11,1 ‘Fetch Output Voltage for 62000B with address 1 Delayms ‘Delay for 62000B responds Can_Rx ‘Receive message End Function 4-16...
Follow the actions described in this chapter to inspect the instrument and troubleshoot the problem first when the 62000B Series DC Power Supply is unable to operate normally. Please consult the sales agent or distributor of Chroma if the information provided in this manual is unable to resolve the problem.
Page 56
Kuei-shan Hwaya Technology Park Taoyuan County 33383, Taiwan 33383 台灣桃園縣龜山鄉 華亞科技園區華亞一路 66 號 T +886-3-327-9999 F +886-3-327-8898 Mail: info@chromaate.com 0Hhttp://www.chromaate.com Copyright by CHROMA ATE INC. All Rights Reserved. All other trade names referenced are the properties of their respective companies.