Chroma 63610-80-20 Operation & Programming Manual page 200

Programmable dc electronic load
Table of Contents

Advertisement

Programmable DC Electronic Load 63600 Series Operation & Programming Manual
MEASure:ALLVoltage?
Type:
Description:
Query Syntax:
Return Parameters:<aard>, [Unit = Volt]
Query Example:
Return Example:
MEASure:CURRent?
Type:
Description:
Query Syntax:
Return Parameters:<NR2>, [Unit = Ampere]
Query Example:
Return Example:
MEASure:INPut
Type:
Description:
Setting Syntax:
Setting Parameters:<NRf>, LOAD(0), UUT(1)
Setting Example:
Query Syntax:
Return Parameters: <CRD>, LOAD, UUT
Query Example:
MEASure:POWer?
Type:
Description:
Query Syntax:
Return Parameters:<NR2>, [Unit = Watt]
Query Example:
Return Example:
MEASure:VOLTage?
Type:
Description:
Query Syntax:
Return Parameters:<NR2>, [Unit = Volt]
Query Example:
Return Example:
5-68
All Channel
Returns real time voltage measured at the input of the all load
channel. The return value is 0 when the channel is not existed.
MEASure:ALLVoltage?
MEAS:ALLV?
1.2, 2, 0, 0, 10.2, 0, 0, 0
Channel-Specific
Returns the real time current measured at the load module input.
MEASure:CURRent?
MEAS:CURR?
3.15
Channel-Specific
Selects the input port of the electronic load to measure voltage.
MEASure:INPut<space><NRf>
MEAS:INP LOAD
MEAS:INP 1
MEASure:INPut?
MEAS:INP?
Channel-Specific
Returns the real time power measured at the load module input.
MEASure:POWer?
MEAS:POW?
3.15
Channel-Specific
Returns the real time voltage measured at load module input.
MEASure:VOLTage?
MEAS:VOLT?
8.12