Chroma 63200 Series Operation & Programming Manual

High power dc electronic load
Hide thumbs Also See for 63200 Series:
Table of Contents

Advertisement

Quick Links

Advertisement

Table of Contents
loading

Summary of Contents for Chroma 63200 Series

  • Page 3 High Power DC Electronic Load 63200 Series Operation & Programming Manual Version 2.9 September 2015 P/N A11 000599...
  • Page 4 The information in this document is subject to change without notice. Chroma ATE INC. makes no warranty of any kind with regard to this manual, including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose.
  • Page 5 All of Chroma’s instruments are warranted against defects in material and workmanship for a period of one year from date of shipment. Chroma agrees to repair or replace any assembly or component found to be defective, under normal use during this period. Chroma’s obligation under this warranty is limited solely to repairing any such instrument, which in Chroma’s sole...
  • Page 6 Material Contents Declaration The recycling label shown on the product indicates the Hazardous Substances contained in the product as the table listed below. : See <Table 1>. : See <Table 2>. <Table 1> Hazardous Substances Lead Mercury Cadmium Hexavalent Polybrominated Polybromodiphenyl Part Name Chromium...
  • Page 7 “” indicates that the level of the specified chemical substance exceeds the threshold level specified in the standards of SJ/T-11363-2006 and EU 2005/618/EC. Chroma is not fully transitioned to lead-free solder assembly at this moment; however, most of the components used are RoHS compliant.
  • Page 10 Failure to comply with these precautions or specific WARNINGS given elsewhere in this manual will violate safety standards of design, manufacture, and intended use of the instrument. Chroma assumes no liability for the customer’s failure to comply with these requirements.
  • Page 11 Safety Symbols DANGER – High voltage. Explanation: To avoid injury, death of personnel, or damage to the instrument, the operator must refer to the explanation in the instruction manual. High temperature: This symbol indicates the temperature is hazardous to human beings. Do not touch it to avoid any personal injury.
  • Page 12 Add a note to the “Specifications” for the fuse of 63205 Jan. 2004 Modify the “Specifications” in Chapter 1 for power rating changes for all models. Add a section of “Dimension Outline for 63200 Series” in Chapter 1. Jan. 2005 Add the following sections: –...
  • Page 13 Minimum setting to 0V in CV Mode − External ON/OFF to support all Modes and Programs − 63202 specifications to upgrade to 600V − Copyright text from “CHROMA ATE Inc.” to “Chroma ATE Inc.” − List of accessories Jun. 2013 Add the following: −...
  • Page 15: Table Of Contents

    High Power DC Electronic Load 63200 Series Operation & Programming Manual Table of Contents General Information ....................1-1 Introduction ..................... 1-1 Description ...................... 1-1 Overview of Key Features ................1-2 Specifications ....................1-2 Dimension Outline for 63200 Series .............. 1-11 Installation ......................
  • Page 16 High Power DC Electronic Load 63200 Series Operation & Programming Manual 4.2.1 Setting the Operation Mode ..............4-3 4.2.2 Setting the Program ................4-9 4.2.3 Running the Program ................4-12 4.2.4 Setting the Specification ................ 4-12 4.2.5 Setting the Configuration ............... 4-13 4.2.6...
  • Page 17 High Power DC Electronic Load 63200 Series Operation & Programming Manual 10.1 Overview ....................... 10-1 10.2 Troubleshooting .................... 10-1 Precautions for Battery Loading ..............11-1 11.1 Measures for Improvement ................11-3 11.1.1 Additional Protection Switch ..............11-3 11.1.2 Operation ....................11-3 Verification ......................
  • Page 19: General Information

    EN61010-1: TYPE POLLUTION II and INSTALLTION CATEGORY II. Description The functions of the 63200 series loads are the same except the variations on input voltage, load current, and power ratings. They can be operated independently in constant current (CC) mode, constant resistance (CR) mode, constant voltage (CV) mode or constant power (CP) mode.
  • Page 20: Overview Of Key Features

    High Power DC Electronic Load 63200 Series Operation & Programming Manual Overview of Key Features  Local operation on front panel keypad  Remote control via GPIB, RS-232 interface or remote controller (optional)  Photo-couple isolation supplies true floating Load ...
  • Page 21 General Information CAT II – is for measurements performed on circuits directly connected to the low-voltage installation. CAT I – is for measurements performed on circuits directly connected to mains. SPECIFICATIONS Model 63201 63202 63203 Power*1 260W 2600W 260W 2600W 520W 5200W Current...
  • Page 22 High Power DC Electronic Load 63200 Series Operation & Programming Manual General Short circuit Current 300A 600A 177 × 440 × 589.4 mm / 177 × 440 × 589.4 mm / 353 × 440 × 589.4 mm / Dimension (H × W × D) 6.96 ×...
  • Page 23 General Information Model 63204 63205 63206 Power*1 520W 5200W 650W 6500W 1040W 10400W Current 0-10A 0-100A 0-18A 0-180A 0-60A 0-600A Voltage 0-600V 0-80V 0-80V 1.25V @ 5A 1.25V @ 50A 0.5V @ 9A 0.5V @ 90A 0.5V @ 30A 0.5V @ 300A Min.
  • Page 24 High Power DC Electronic Load 63200 Series Operation & Programming Manual General Short circuit Current 100A 180A 600A 353 × 440 × 589.4 mm / 310 × 440 × 589.4 mm / 443.7 × 440 × 589.4 mm / Dimension (H × W × D) 13.89 ×...
  • Page 25 General Information Model 63207 63208 63209 Power*1 1040W 10400W 1560W 15600W 1560W 15600W Current 0-30A 0-300A 0-60A 0-600A 0-100A 0-1000A Voltage 0-80V 0-80V 0-80V 0.5V @ 15A 0.5V @ 150A 0.5V @ 30A 0.5V @ 300A 0.5V @ 50A 0.5V @ 500A Min.
  • Page 26 High Power DC Electronic Load 63200 Series Operation & Programming Manual General Short circuit Current 300A 600A 100A 1000A 443.7 × 440 × 589.4 mm / 762.8 × 546 × 700 mm / 762.8 × 546 × 700 mm / Dimension (H ×...
  • Page 27 General Information Model 63210 63211 63212 Power*1 1450W 14500W 15600W 10000W Current 0-15A 0-150A 0-30A 0-150A 0-30A 0-150A Voltage 0-600V 10-1000V 10-1000V 1.5V @ 7.5A 1.5V @ 75A 5V @ 15A 5V @ 75A 5V @ 15A 5V @ 75A Min.
  • Page 28 High Power DC Electronic Load 63200 Series Operation & Programming Manual General Short circuit Current 150A 150A 150A 762.8 × 546 × 700mm / 762.8×546×700mm / 762.8×546×700mm / Dimension (H × W × D) 30.03×21.49×27.55inch 30×21.5×27.6inch (cabinet) 30×21.5×27.6inch (cabinet) (cabinet) Weight 170 kg / 374.78 lbs...
  • Page 29: Dimension Outline For 63200 Series

    General Information Dimension Outline for 63200 Series  Model 63201 and 63202  Model 63203 and 63204 1-11...
  • Page 30 High Power DC Electronic Load 63200 Series Operation & Programming Manual  Model 63205  Model 63206 and 63207 1-12...
  • Page 31 General Information  Model 63208, 63209, 63210 (Top View) Top Side Common Difference ± 2.0 mm  Model 63211, 63212 Unit: mm 1-13...
  • Page 32 High Power DC Electronic Load 63200 Series Operation & Programming Manual Rack mount note for model 63207: Notices for Rack Mount Installation 1-14...
  • Page 33 General Information Installing the 63200 Protective Cover: Step 1: Insert the cable into protective cover. Step 2: Socket the Bakelite into the positive/negative copper plate until it touches the chassis and then secure the M6x20L screw and M6 screw nut to block the Bakelite.
  • Page 35: Installation

    Installation Installation Introduction This chapter describes how to install the 63200 series Load as well as a turn-on check procedure and application considerations. Inspection As soon as the instrument is unpacked, inspect any damage that might have occurred in shipping. Keep all packing materials in case that the instrument has to be returned. If any damage is found, please file a claim to the carrier immediately.
  • Page 36 High Power DC Electronic Load 63200 Series Operation & Programming Manual 27 Power Cord 110V 28 Fuse (5A/250V) Standard Accessories Item Quick Start Guide Quick Start Guide Name Fuse(2A/250V) Fuse (2.5A/250V) T-Chinese Version English Version Item Installing kit for Name Fuse (3.15A/250V)
  • Page 37: Maintenance & Cleaning

    Be sure to verify the accuracy of the Electronic Load as described in Chapter 12 Verification every 6 months. If there is service request for the 63200 or the specification does not match, please contact the global sales and service location listed on the Chroma’s web page below: http://www.chromaate.com/english/contact/default.asp Installing The Electric Loads can operate well within the temperature range from 0 ºC to 40 ºC.
  • Page 38: Turn-On Self-Test

    All of the 7-segment LEDs which display the measuring V, I, P (or R) will be activated when the self-test completes. If any error is found in self-test, the display will stop here. In case of failure, return the Load to Chroma sales or service office for repair.
  • Page 39: Application Connection

    Installation Application Connection 2.4.1 Load Connections WARNING To satisfy safety requirements, load wires must be heavy enough not to overheat while carrying the short-circuit output current of the device connected to the Electronic Load. Input connections are made to the + and − terminal block on the rear of the Load. The major considerations in making input connections are the wire size, length and polarity.
  • Page 40: Remote Sensing Connections

    Connect the negative ends is suggested to avoid any possible loop current that may affect the measurement result. Contact Chroma technical service center if there is a need to use the UUT voltage measurement function as a power meter.
  • Page 41: Parallel Connections

    RS485 ADDRESS and control the parallel load. There is only one Master in the parallel group, the rest are slaves. For 63200 Series, it can parallel up to 6 models and these Electronic Loads can be operated simultaneously via Master in static or dynamic mode.
  • Page 42: Parallel Connection Diagram

    High Power DC Electronic Load 63200 Series Operation & Programming Manual 2.4.4 RS-485 Parallel Connection Diagram 2.4.4.1 Pin Assignment 2.4.4.2 Master & Salve Connecting Diagram...
  • Page 43: Parallel Setting

    FRONT PANEL TESTING [ RS232 ] PASS FRONT PANEL TESTING CHECKSUM PASS GPIB ADDRESS CHROMA ATE INC. MODEL 632XX Next, it will show the CC mode settings. CCHA: 0.0000A CCHB: 0.0000A Press SYS to set the RS485 address of Slave so that it can be followed when setting the attribute of Master.
  • Page 44 High Power DC Electronic Load 63200 Series Operation & Programming Manual SLAVE1 OFFLINE Set the Master as follows: Connect the A port of RS485 cable to the selected Master. The start screen is same as the CC mode settings. CCHA: 0.0000A...
  • Page 45: Remote Control Connection

    These connectors on the rear panel connect the Load to the controller or computer. The GPIB interface of the electronic load is standard. The remote controller is optional. The 63200 series Remote Controller can control the load via RS-232 port. Connect the Remote Controller to the Electronic Load before powering it on.
  • Page 47: Operation Overview

    Operation Overview Introduction Chroma 63200 series electronic loads are suitable for design, manufacturing, testing and quality assurance. It contains a processor, GPIB and RS-232 connectors, front panel keypad, display, and power stage. Its built-in remote control function allows you to control, read back current, voltage and status.
  • Page 48 High Power DC Electronic Load 63200 Series Operation & Programming Manual Figure 3-2 Rear Panel of Model 63201/63202/63203/63204...
  • Page 49 Operation Overview Figure 3-3 Rear Panel of Model 63205/63206/63207...
  • Page 50 High Power DC Electronic Load 63200 Series Operation & Programming Manual Figure 3-4 Rear Panel of Model 63208/63209/63210...
  • Page 51 Operation Overview Figure 3-5 Rear Panel of Model 63211/63212...
  • Page 52: Local/Remote Control

    High Power DC Electronic Load 63200 Series Operation & Programming Manual Local/Remote Control Local (front panel) control is in effect immediately after the power is applied. The front panel keypad and display allow manual control when Load is used in bench test applications.
  • Page 53 Operation Overview In CC mode, the Load will sink a current in accordance with the programmed value regardless of the input voltage. The CC mode can be set by CC key in the front panel. Press A/B key to select the current waveform in Static mode. Static function checks the stability of output voltage from a power supply.
  • Page 54: Constant Resistance Mode

    High Power DC Electronic Load 63200 Series Operation & Programming Manual Current Load1 load2 Time 10mS 10mS 10mS 10mS Figure 3-8 Dynamic Current Waveform Slew Rate (Rise, Fall A/µs) Slew rate determines the rate at which the current input of changes to a newly programmed value.
  • Page 55: Constant Voltage Mode

    Operation Overview Resistance can be programmed in either of low range or high range by the RANGE key. The low range is used for input voltage in low voltage range while the high range is for the input voltage over low voltage range. The current range of CR mode is high range. There are two resistance levels (A or B) for CR function as static CC mode.
  • Page 56: Constant Power Mode

    3.5.5 Load Surge Capability Chroma’s 63200 Series DC Loads provide a unique load surge simulation capability, which allows users to overdrive the loads up to 2.7 times their rated power for short periods. This feature is ideal when the average power require by the UUT is low compared to short-term peak power demands.
  • Page 57 Operation Overview Figure 3-12 Load Surge Capability (Static Mode) Figure 3-13 Load Surge Capability (Dynamic Mode) Note The Initial state under Static Mode should last at least 1 second. This load surge capability will be regulated by the temperature de-rating characteristics.
  • Page 58: Timer Function For Battery Discharge Testing

    High Power DC Electronic Load 63200 Series Operation & Programming Manual Example 2: DYNAMIC LOADING The Model 63201 is capable of a zero – to- 6500W (250%) pulse at a duty cycle of 5%. This is represented by the DOT on the purple curve in Figure 3-13.
  • Page 59: Slew Rate & Minimum Transient Time

    Operation Overview Slew Rate & Minimum Transient Time Slew rate is defined as the change in current over time. A programmable slew rate allows a controlled transition from one load setting to another to minimize induced voltage drops on inductive power wiring, or to control induced transients on a test device. If the transient from one setting to another is large, the actual transient time can be calculated by dividing the current transition by the slew rate.
  • Page 60: Short On/Off

    Please be noted that turning on the short circuit may cause loading too much current to trig protection circuit to turn off the Load. The 63200 series have a TTL signal from RS-485 pin 6 for your application to control the external short relay. TTL indicates high/low when the short key is set on/off.
  • Page 61: Load On/Off

    Operation Overview 3.10 Load On/Off A module’s input can be toggled on/off through the blue LOAD ON/OFF key on the front panel, or the remote control. The on/off change for input is done according to the slew rate. Turning off the load does not affect the programmed setting. The load will return to the previous programmed values when the Load is turned on again.
  • Page 62: Save/Recall Setting

    High Power DC Electronic Load 63200 Series Operation & Programming Manual  Over temperature The Load has an over temperature protection circuit, which will turn off the load if internal temperature exceeds the safe limit. The over temperature status register bit is set when the OT condition occurs, and will remain set till it is reset.
  • Page 63: Program

    3.15 Voltage & Current Monitor The 63200 series have two isolated BNC connectors to monitor load voltage and current, the output signal from I MON and V MON. They are on the rear panel. A signal that maps to full scale will output from BNC connector.
  • Page 64 High Power DC Electronic Load 63200 Series Operation & Programming Manual Example: DC Load Figure 3-17 Connection of Power, UUT and DC Load (1) When it is “LOAD ON”, Von Point is set to 0V and Von Protection is disabled, the DC LOAD will occur overshoot when the SW is turned off.
  • Page 65 Operation Overview (2) When it is “LOAD ON”, Von Point is set to 0V and Von Protection is enabled, the DC Load starts loading when the external voltage exceeds the protection voltage. No overshoot will occur under this condition as Figure 3-19 shows. Output Load Sink Current...
  • Page 66: Cr Offset Error Compensation

    CR Compensation: (1) Advantage: The measurement in CR mode will be more accurate because of compensation and the accuracy of 63200 Series is based on the compensated result. (2) Disadvantage: Controlling current is the mechanism of compensation and in the process of continuous compensation;...
  • Page 67: External Load On/Off Control

    Operation Overview 3.18 External LOAD ON/OFF Control The External Load ON/OFF Control is to enable the DC Load ON/OFF function by external signals via RS-485 cable. It can support all modes (Ex: CC, CR, CP, CV & Program.) First, set the PARALLEL in SYS to NONE. PARALLEL = NONE PRESS 0→3 TO SELECT Next, set the EXT.
  • Page 68: Short Key:enable/Disable Function

    High Power DC Electronic Load 63200 Series Operation & Programming Manual Timing Chart: 輸入訊號 Input Signal Load 100ms 100ms 100ms 100ms 3.19 SHORT KEY:ENABLE/DISABLE Function This function allows the user to set the SHORT key to be valid or invalid. The SHORT key is valid for execution short function when enabled and is invalid for executing short function when disabled.
  • Page 69: Local Operation

    Local Operation Local Operation Introduction This chapter describes how to operate the electronic load from the local panel in details. The descriptions include panel control and indicators. Local Operation In order to use the front panel keys to control the electronic load, local operation must be in effect.
  • Page 70 High Power DC Electronic Load 63200 Series Operation & Programming Manual Line switch Turn the ac power on/off. LCD display Display setting information normally. 7-segment LEDs Show the V, I, P, R measured data. System keys To set the remote control status, ext. wave enable and Parallel function parameters.
  • Page 71: Setting The Operation Mode

    Local Operation Entry keys ▲, ▼ They let you scroll through the choices in a parameter list for a specific command. Parameter lists are circular. You can return to the starting position by pressing either key continuously. CLEAR To clear the digit entered from keypad. This key lets you correct wrong digits before they are entered.
  • Page 72 High Power DC Electronic Load 63200 Series Operation & Programming Manual Set Current Level There are 4000 discrete steps from 0 to full scale in each range. Set level A current level to 200A by pressing 2, 0, 0, ENTER. Set level B current level to 5A by pressing 5, ENTER.
  • Page 73 Local Operation CCDLT1: 0.100ms CCDLT2: 0.200ms The slew rate and Voltage Spec settings are same as those of static current mode. If you press ENTER key, and the blinked data do not go to next, change the configuration setting Enter Data Next to YES. 4.2.1.2 Setting CR Values The Load CR mode is programmed by pressing CR.
  • Page 74 High Power DC Electronic Load 63200 Series Operation & Programming Manual CRH VOLTAGE SPEC. LOW : 4.000V Then the display will go to the first editing page again. 4.2.1.3 Setting CV Values The CV mode for the Load is programmed by pressing CV. The voltage values can be programmed in low voltage (CVL) or high voltage (CVH) range.
  • Page 75 Local Operation Figure 4-2 CV Response Transfer Function (FAST) Figure 4-3 CV Response Transfer Function (SLOW) Set Current spec Set the high current spec. to 8 A by pressing 8, ENTER. CVL CURRENT SPEC. HIGH : 8.000A Set the low current spec. to 4 A by pressing 4, ENTER. CVL CURRENT SPEC.
  • Page 76 High Power DC Electronic Load 63200 Series Operation & Programming Manual CC, CP V RANGE SELECT 1: HIGH 2: LOW Press ENTER to the selections on next page and set Von PROTECTION to NO by selecting 2. Von PROTECTION 1:YES 2:NO Press ENTER to the selections on next page and set Von POINT to 0V by selecting 0.
  • Page 77: Setting The Program

    Local Operation Set Power Level Set level A power level to 200 W by pressing 2, 0, 0, ENTER. Set level B power level to 4000 W by pressing 4, 0, 0, 0, ENTER. CPHA : 200.000W CPHB : 4000.000W Set Slew Rate Set the rise 20 A/µs and fall slew rates to 0.5A/µs by pressing 2, 0, ENTER for rise and 0, .
  • Page 78 High Power DC Electronic Load 63200 Series Operation & Programming Manual Table 4-1 The Relationship of the Program Sequence and the Corresponding File Program 1 Sequence No. Corresponding File No. Program 2 Sequence No. Corresponding File No. Program 10 Sequence No.
  • Page 79 Local Operation Setting the Sequence Load ON/OFF Time The sequence ON/OFF time controls the Load input ON/OFF when the program’s sequence is executed. The ON/OFF time range is 0 to 60 seconds. (30s each) SEQUENCE 1 LOAD ON TIME : 20.0 s Press 8, ENTER keys to set OFF time for 8 seconds.
  • Page 80: Running The Program

    High Power DC Electronic Load 63200 Series Operation & Programming Manual 4.2.3 Running the Program Press LOAD ON/OFF to run program when the program function is selected. The display shows as below. PROG. 1 SEQ. 1 [ON ][ KEY ] [ PASS ] [ S ] The upper line shows the executed program and sequence number while the lower line shows the Load, key, test result and short status.
  • Page 81: Setting The Configuration

    Local Operation 4.2.5 Setting the Configuration The Electronic Load provides useful features such as Von point, Current limit, Auto Load On, etc. To use these powerful features, you must set relevant parameters in accordance with application needs by the configuration setup. This procedure is only needed for initial setup of a test operation.
  • Page 82 High Power DC Electronic Load 63200 Series Operation & Programming Manual CAUTION If Von is set to 0V, the Load circuit will be ON in spite of no UUT. This will get overshoot spike. If a UUT is applied, the overshoot may damage the UUT even though the Load current sets small.
  • Page 83 Local Operation Set CV mode CURR_LIMIT. This function will limit the current loading to protect the UUT in CV mode. The default setting of current limit is the maximum Load current. CV CURRENT LIMIT CURRENT = 200.000A Set sign of voltage for display. The Electronic Load will show minus sign of the voltage if MINUS is selected.
  • Page 84 High Power DC Electronic Load 63200 Series Operation & Programming Manual Select the load on mode if AUTO LOAD ON is ON. If LOAD is selected, the Load will be active as DEFAULT setting. If PROG is selected, the Load will be active as the program saved last time.
  • Page 85: Recalling Files

    Local Operation Display the versions of host, load, and Panel. <HOST FW> WRTR: 02.00 BOOT: 02.00 MAIN: 02.31 Press ▼ key and the display will show: <LOAD FW> WRTR: 02.00 BOOT: 02.00 MAIN: 02.31 Press ▼ key and the display will show: <PANEL FW>...
  • Page 86: Going To Local

    High Power DC Electronic Load 63200 Series Operation & Programming Manual Press SAVE, ▲, ▼ until the display shows as follows. Press 1, ENTER to save program. SAVE PROGRAM 1: YES 2: NO 4.2.8 Going To Local The CONF./LOCAL key operates as local key when Electronic Load is in remote mode. You can press CONF./LOCAL key to go to local operation when Load is in remote state.
  • Page 87: Online Change Level

    Local Operation Note Pin 1 and pin 9 (+5V) are for 63200 series Remote Controller only. External wave Enable. The Load current can be controlled by the external DC voltage input via the Vext BNC connector on the rear panel if this function is selected 1: YES.
  • Page 89: Basic Information For Programming

    Basic Information for Programming Basic Information for Programming Introduction The following sections describe how to program the 63200 series electronic load remotely from a GPIB controller or RS-232. The command set introduced here can be applied to all 63200 series electronic loads.
  • Page 90: Gpib Capability Of The Electronic Load

    High Power DC Electronic Load 63200 Series Operation & Programming Manual Electronic Load IBM PC No Connection No Connection No Connection No Connection No Connection No Connection GPIB Capability of the Electronic Load Table 5-2 GPIB Capability Response Interface Functions All electronic load functions are programmable over the GPIB.
  • Page 91: In Remote Control

    Basic Information for Programming RS-232 in Remote Control When you use RS-232 in remote control, you have to send the remote command CONFigure : REMote ON first to let control procedure enter into remote state, and then execute other command set. When control comes to an end, you have to send the command CONFigure : REMote OFF to let control procedure return to local operation mode.
  • Page 93: Introduction To Programming

    Introduction to Programming Introduction to Programming Basic Definition GPIB statement includes instrument control and query commands. A command statement sends an instruction to the electronic load and query command request information from the electronic load. Simple Command A simple command statement consists of a command or keyword usually followed by a parameter or data: LOAD ON or LOAD:SHORT OFF...
  • Page 94: Numerical Data Formats

    Chroma 63200 electronic load accepts the numerical data type listed in Table 6-1. Numeric data may be followed by a suffix to specify the dimension of the data. A suffix may be preceded by a multiplier. Chroma 63200 makes use of the suffixes listed in Table 6-2 and multipliers listed in Table 6-3.
  • Page 95: Separators And Terminators

    Introduction to Programming Separators and Terminators In addition to keywords and parameters, GPIB program statements require the following: Data Separators: Data must be separated from the previous command keyword by a space. This is shown in examples as a space (CURR 3) and on diagrams by the letters SP inside a circle. Keyword Separators: Keywords (or headers) are separated by colon (:), semicolon (;), or both.
  • Page 96 High Power DC Electronic Load 63200 Series Operation & Programming Manual Please refer to the following figure. (root):VOLT:L1: 30<nl> Starting a New Line to return to the Root. (root):SPEC:VOLT:H 30; :L 5;: (root)RES:L1 400; :RISE 1000;:...
  • Page 97: Language Dictionary

    Language Dictionary Language Dictionary Commands for operating the 63200 Electronic Load remotely are grouped into subsystems. Each command belonging to the same subsystem is arranged in alphabetic order. A syntax chart of the subsystem that contains the commands in the same group is included. Sub- systems are ordered alphabetically according to their names in the following sections.
  • Page 98 High Power DC Electronic Load 63200 Series Operation & Programming Manual Parameter 0 to 255 Example *ESE 48 This command enables the CME and EXE events for the Standard Event Status Event register. Query Syntax *ESE? Return Parameter <NR1> Query Example *ESE? This query returns the current setting for "Standard...
  • Page 99 Language Dictionary *RCL Recall Instrument State Command Description This command restores the electronic load to a state that was previously stored in memory with *SAV command to the specified location (see *SAV). Syntax *RCL <NRf> Parameter 1 to 101, 101 returns to factory default. Example *RCL 50 *RST...
  • Page 100: Specific Commands

    QUES = questionable status summary MAV = message available Query Syntax *STB? Return Parameter <NR1> Query Example *STB? Return the contents of "Status Byte". Return Example Specific Commands The 63200 series products are built-in with the following specific GPIB commands.
  • Page 101: Configure Sub-System

    Language Dictionary 7.2.1 CONFigure Sub-system CONFigure :AUTO :LOAD :MODE :DISPlay :KEY :REMote :SAVE :SOUNd :VOLTage :LATCh :RESet :POLarity :RANGe :PROTection :BATT :VOLT :Capacity :TIME :TIMEOUT :CROFFSET :SHORt :KEY CONFigure:AUTO:LOAD Description Set if the load module will do Auto Load On during power-on. Syntax CONFigure:AUTO:LOAD <NR1 | CHAR>...
  • Page 102 High Power DC Electronic Load 63200 Series Operation & Programming Manual CONFigure:DISPlay Description Set measure display mode Command Syntax CONFigure:DISPlay <NR1 | CHAR> Parameter 0 | RESISTANCE, 1 | POWER Example CONF:DISP POWER Set the power display mode. CONF:DISP 0 Set resistance display mode.
  • Page 103 Language Dictionary Return Parameter 0 | 1 Query Example CONF:VOLT: LATC? Return the action type to Von. Return Example 0 (OFF) CONFigure:VOLTage:LATCh:RESet Description Reset Von signal. Syntax CONFigure:VOLTage:LATCh:RESet Parameter None Example CONF:VOLT:LATC:RES CONFigure:VOLTage:ON Description Set the voltage of loading current to on. Syntax CONFigure:VOLTage:ON <NRf>...
  • Page 104 High Power DC Electronic Load 63200 Series Operation & Programming Manual CONFigure:BATT Description Set the battery discharge measurement. Syntax CONFigure:BATT <NR1 | CHAR> Parameter 0 | OFF, 1 | ON Example CONF:BATT 1 Enable battery discharge. CONF:BATT OFF Disable battery discharge.
  • Page 105 Language Dictionary CONFigure:CROFFSET Description Set compensation mechanism in CR mode. Syntax CONFigure:CROFFSET <NR1 | CHAR> Parameter 0 | OFF, 1 | ON Example CONF:CROFFSET 1 Enable CR compensation. CONF:CROFFSET OFF Disable CR compensation. Query Syntax CONFigure:CROFFSET? Return Parameter 0 | 1 Query Example CONF:CROFFSET? Return CR compensation state.
  • Page 106: Communicate Sub-System

    High Power DC Electronic Load 63200 Series Operation & Programming Manual 7.2.2 COMMunicate Sub-system COMMunicate :ADDRess :GPIB :RS485 :SERial :BAUD :BITS :PARity :SBITs COMMunicate:ADDRess:GPIB Description Set the GPIB address Syntax COMMunicate:ADDRess:GPIB <NR1> Parameter 1 – 30 Example COMM:ADDR:GPIB 16 Query Syntax...
  • Page 107 Language Dictionary Return Example 0 ( BITS_8) COMMunicate:SERial:PARity Description Set the parity scheme that is to be used for transmission Syntax COMMunicate:SERial:PARity <NR1 | CHAR> Parameter 0 | NONE, 1 | EVEN, 2 | ODD Example COMM: SER:PAR 0 COMM: SER:PAR NONE Query Syntax COMMunicate:SERial:PARity? Return Parameter...
  • Page 108: Current Sub-System

    High Power DC Electronic Load 63200 Series Operation & Programming Manual 7.2.3 CURRENT Sub-system CURRent :STATic :RISE :FALL :DYNamic :RISE :FALL CURRent:STATic Description Set A or B Static in CC mode for load. Syntax CURRent:STATic <NR1 | CHAR> Parameter 0 | B, 1 | A...
  • Page 109 Language Dictionary CURRent:STATic:RISE/FALL Description Set the current slew rate for constant current static mode. Syntax CURRent:STATic:RISE <NRf+>[suffix] CURRent:STATic:FALL <NRf+>[suffix] Parameter For valid value range refer to respective specification. Example CURR:STAT:RISE 2.5 Set rise slew rate to 2.5A/µs for static load. CURR:STAT:FALL 1A/µs Set fall slew rate to 1A/µs for static load.
  • Page 110 High Power DC Electronic Load 63200 Series Operation & Programming Manual Query Example CURR:DYN:RISE? Return the rise slew rate of dynamic load. Return Example CURRent:DYNamic:T1/T2 Description Set the duration parameter T1 or T2 for dynamic load. Syntax CURRent:DYNamic:T1 <NRf+>[suffix] CURRent:DYNamic:T2 <NRf+>[suffix] Parameter For valid value range refer to respective specification.
  • Page 111: Fetch Sub-System

    Language Dictionary 7.2.4 FETCh Sub-system FETCh :CURRent :POWer :RESistance :VOLTAGE :STATus FETCh:CURRent? Description Return the current measured at the input of the load. Query Syntax FETCh:CURRent? Return Parameter <NR2> [Unit=Ampere] Query Example FETC:CURR? Return Example 3.15 FETCh:POWer? Description Return the power measured at the input of the load. Query Syntax FETCh:POWer? Return Parameter...
  • Page 112: Load Sub-System

    High Power DC Electronic Load 63200 Series Operation & Programming Manual 7.2.5 LOAD Sub-system LOAD [:STATe] :SHORt [:STATe] :KEY :PROTection :CLEar :SAVE LOAD[:STATe] Description The LOAD command makes the electronic load active on or off. Syntax LOAD:[STATe] <NR1 | CHAR >...
  • Page 113 Language Dictionary LOAD:PROTection? Description This command returns the status of the electronic load. Query Syntax LOAD:PROTection? Return Parameter <NR1> Bit Position 15 14 13 12 11 10 Condition 0 OT RV OP OV OC Bit Weight Query Example LOAD:PROT? Return the status of the electronic load. Return Example LOAD:PROTection:CLEar Description...
  • Page 114: Measure Sub-System

    High Power DC Electronic Load 63200 Series Operation & Programming Manual 7.2.6 MEASure Sub-system MEASure :CURRent :INPut :POWer :RESistance :STATus :VOLTage MEASure:CURRent? Description Return the real time current measured at the input of the electronic load. Query Syntax MEASure:CURRent? Return Parameter <NR2>...
  • Page 115 Language Dictionary MEASure:POWer? Description Return the real time power measured at the input of the electronic load. Query Syntax MEASure:POWer? Return Parameter <NR2> [Unit=Watt] Query Example MEAS:POW? Return Example 8.12 MEASure:VOLTage? Description Return the real time voltage measured at the input of the electronic load.
  • Page 116: Mode Sub-System

    High Power DC Electronic Load 63200 Series Operation & Programming Manual 7.2.7 MODE Sub-system MODE MODE Description This command sets operational modes of the electronic load. Syntax MODE <NR1 | CHAR> Parameter 0 | CCL, 1 | CCH, 2 | CCDL, 3 | CCDH, 4 | CRL, 5 | CRH, 6 | CVL, 7 | CVH, 8 | CPL, 9 | CPH, 10 | CCEL, 11 | CCEH.
  • Page 117: Power Sub-System

    Language Dictionary 7.2.8 POWer Sub-system POWer :RISE :FALL POWer Description Set A or B Static in CP mode for load. Syntax POWer <NR1 | CHAR> Parameter 0 | B, 1 | A Example POW 0 Set CP mode to Static B. POWer:L1/L2 Description Set the Static Load Current of constant power mode.
  • Page 118: Program Sub-System

    High Power DC Electronic Load 63200 Series Operation & Programming Manual 7.2.9 PROGram Sub-system PROGram :CHAin :FILE :KEY :ONTime :OFFTime :DELay :TIME :RUN :SAVE :SEQuence :MODE :SHORt :DELay :TIME PROGram:CHAin Description Set the type of program file in serial execution.
  • Page 119 Language Dictionary Query Example PROG:ONT? Return Example PROGram:OFFTime Description Set the load off time of program file Syntax PROGram:OFFTime <NRf>[suffix] Parameter 0~30 Sec Example PROG:OFFT 20 PROG:OFFT 200ms Query Syntax PROGram:OFFTime? Return Parameter <NR2> [Unit=Sec] Query Example PROG:OFFT? Return Example PROGram:PF:DELay Description Set the pass/fail measurement delay time of program file.
  • Page 120 High Power DC Electronic Load 63200 Series Operation & Programming Manual Syntax PROGram:SEQuence <NRf> Parameter 1 to 10 Example PROG:SEQ 3 Query Syntax PROGram:SEQuence? Return Parameter <NR1> Query Example PROG:SEQ? Return Example PROGram:SEQuence:MODE Description Set the sequence type. Syntax PROGram:SEQuence:MODE <NR1 | CHAR>...
  • Page 121: Resistance Sub-System

    Language Dictionary 7.2.10 RESistance Sub-system RESistance :RISE :FALL RESistance Description Set A or B Static in CR mode for load. Syntax RESistance <NR1 | CHAR> Parameter 0 | B, 1 | A Example RES 0 Set CR mode to Static B. RESistance:L1/L2 Description Set the static resistance level for constant resistance mode.
  • Page 122: Specification Sub-System

    High Power DC Electronic Load 63200 Series Operation & Programming Manual 7.2.11 SPECification Sub-system SPECification [:PASS] :TEST :UNIT SPECification:C/L/H Description Set the voltage (CC, CR, CP mode) or current (CV mode) specification. Syntax SPECification:H SPECification:L SPECification:C Parameter For valid value range refer to respective specification.
  • Page 123 Language Dictionary Return Parameter 0 | 1 Return Example 0 (PERCENT) 7-27...
  • Page 124: Status Sub-System

    High Power DC Electronic Load 63200 Series Operation & Programming Manual 7.2.12 STATus Sub-system STATus :QUEStionable :CONDition :ENABle [:EVENt] :NTRansition :PTRansition STATus:QUEStionable:CONDition Description Real-time ("live") recording for Questionable data Query Syntax STATus:QUEStionable:CONDition? Return Parameter <NR1> Query Example STAT:QUES:COND? Return the channel status.
  • Page 125 Language Dictionary STAT:QUES:NTR 4 Set OP(over power bit 2) to 1-to-0. Query Syntax STATus:QUEStionable:PTRansition? STATus:QUEStionable:NTRansition? Return Parameter <NR1> Query Example STAT:QUES:PTR? Return the setting on the QUEStionable Ptransition/Ntransition. Return Example 7-29...
  • Page 126: Voltage Sub-System

    High Power DC Electronic Load 63200 Series Operation & Programming Manual 7.2.13 VOLTage Sub-system VOLTage :CURRent :MODE VOLTage Description Set A or B Static in CV mode for load. Syntax VOLTage <NR1 | CHAR> Parameter 0 | B, 1 | A...
  • Page 127 Language Dictionary Return Parameter 0 | 1 Query Example VOLT:MODE? Return Example 0 (SLOW) 7-31...
  • Page 128: System Sub-System

    High Power DC Electronic Load 63200 Series Operation & Programming Manual 7.2.14 SYSTem Sub-system SYSTem :WAVE :PARallel :DATA :DISPlay :MODE :SLAVe :CHANnel :MODEl SYSTem:WAVE Description Set the waveform source. Syntax SYSTem:WAVE <NR1 | CHAR> Parameter 0 | INTERNAL, 1 | EXTERNAL...
  • Page 129 Language Dictionary SYSTem:PARallel:SLAVe:CHANnel Description Set parallel mode for slave channel. Syntax SYSTem:PARallel:SLAVe:CHANnel <NR1> Parameter Example SYST:PAR:SLAV:CHAN 1 Query Syntax SYSTem:PARallel:SLAVe:CHANnel? Return Parameter <NR1> Query Example SYST:PAR:SLAV:CHAN? Return Example SYSTem:PARallel:SLAVe:MODEl Description Set parallel for slave model. Syntax SYSTem:PARallel:SLAVe:MODEl <NR1 | CHAR> Parameter (For Model63201~63210) 0 | NONE...
  • Page 131: Status Reporting

    Status Reporting Introduction This chapter covers the status data structure of Chroma 63200 series electronic load as shown in Figure 8-1 (on the next page). The standard registers, such as the Event Status register group, the Output Queue, the Status Byte and Service Request Enable registers perform standard GPIB functions and are defined in IEEE-488.2 Standard Digital Interface for...
  • Page 132: Questionable Status

    High Power DC Electronic Load 63200 Series Operation & Programming Manual QUESTIONABLE STATUS CONDITION PTR/NTR EVENT ENABLE N.U. 5-15 SERVICE REQUEST GENERATION OUTPUT QUEUE STANDARD EVENT STATUS SERVICE STATUS REQUEST BYTE 256 BYTE GENERATION EVENT ENABLE DATA N.U. DATA N.U.
  • Page 133: Output Queue

    Status Reporting Table 8-1 Bit Description of Questionable Status Mnemonic Bit Value Meaning Over current. When an over current condition has occurred on a channel, Bit 0 is set and remains set until the over current condition is removed and LOAD:PROT:CLE is programmed. Over voltage.
  • Page 134: Standard Event Status

    High Power DC Electronic Load 63200 Series Operation & Programming Manual Standard Event Status All programming errors that have occurred will set one or more of the error bits in the Standard Event Status register. ■ Table 8-2 describes the standard events that apply to the electronic load.
  • Page 135: Service Request Enable Register

    Status Reporting Service Request Enable Register ■ The Service Request Enable register can be programmed to specify the bit in the Status Byte register that will generate service requests.
  • Page 137: Command Index

    Command Index Command Index Command Summary Common Commands *CLS *ESE <NR1> *ESE? *ESR? *IDN? *OPC *OPC? *RCL <NRf> *RST *SAV <NRf> *SRE <NR1> *SRE? *STB? Instrument Commands CONFigure :AUTO :LOAD :MODE :DISPlay :KEY :REMote :SAVE :SOUNd :VOLTage :LATCh :RESet :POLarity :RANGe :PROTection :BATT...
  • Page 138 High Power DC Electronic Load 63200 Series Operation & Programming Manual COMMunicate :ADDRess :GPIB :RS485 :SERial :BAUD :BITS :PARity :SBITs CURRent :STATic :RISE :FALL :DYNamic :RISE :FALL FETCh :CURRent :POWer :RESistance :VOLTAGE :STATus LOAD [:STATe] :SHORt [:STATe] :KEY :PROTection :CLEar...
  • Page 139 Command Index :FALL PROGram :CHAin :FILE :KEY :ONTime :OFFTime :DELay :TIME :RUN :SAVE :SEQuence :MODE :SHORt :DLEay :TIME RESistance :RISE :FALL SPECification [:PASS] :TEST :UNIT STATus :QUEStionable :CONDition :ENABle [:EVENt] :NTRansition :PTRansition VOLTage :CURRent :MODE SYSTem :WAVE :PARallel :DATA :DISPlay :MODE :SLAVe...
  • Page 140: Parallel Operation

    High Power DC Electronic Load 63200 Series Operation & Programming Manual :CHANnel :MODEl Parallel Operation GPIB COMMAND MASTER SETTING SLAVE SETTING GPIB COMMAND PARALLEL MODE PARALLEL MODE SYSTem:PARallel:MODE SYSTem:PARallel:MODE SETTING SETTING SLAVE MODEL RS485 ADDRESS SYSTem:PARallel:SLAVe:CHANnel X COMMunication:ADDRess:RS485 SYSTem:PARallel:SLAVe:MODEl X...
  • Page 141: Troubleshooting

    If the 632XX Series DC Electronic Load is unable to operate normally, follow the descriptions in this chapter below to troubleshoot the problems first. Please consult with Chroma’s agents or distributors if the information provided in the manual cannot solve the problem.
  • Page 143: Precautions For Battery Loading

    In regard of the blooming EV, the test application for high power battery has become more and more; and the 63200 Series high power, high voltage electronics loads such as 63204 and 63210 that comply to customer’s the most. However, since it is to test the battery with high power and voltage, it is necessary to pay more attention to the application safety.
  • Page 144 High Power DC Electronic Load 63200 Series Operation & Programming Manual Ls 10u R1 0 Cs 10u V1 500 R2 100m Figure 11-2 Simulated Circuit Diagram 1.00k VF1=911V VF1=911V 750.00 500.00 250.00 0.00 0.00 250.00u 500.00u 750.00u 1.00m 時間(s ) Time(s) Figure 11-3 Simulation of Surge Chart when Switching between Electronic Load &...
  • Page 145: Measures For Improvement

    Precautions for Battery Loading 11.1 Measures for Improvement 11.1.1 Additional Protection Switch As the burnout may expand due to the MOSFET damage and continuous energy release from battery that caused by the conditions described previously, it is suggested to connect the wires as Figure 11-4 shows below when doing the battery charge/discharge tests to prevent problems from happening and to ensure the safety of using Electronic Load.
  • Page 146 High Power DC Electronic Load 63200 Series Operation & Programming Manual  When NFB is selected, since the battery maximum current is 100A, the NFB should be smaller than 100A; therefore it is suggested to use NFB of 20A. ...
  • Page 147: Verification

    Electronic Loads and testing the specifications. The testing is performed using the Chroma 63200 Series models and some necessary devices as listed in Table 12-1. Please see 12.3 Performance Test for connecting the hardware and the test procedure. The user can use the verification checklist to record the specification verified.
  • Page 148 Select a proper range for the current shunt resistance. Press CC and RANGE on the front panel of Chroma 63200 to select the CC high range and program current as Table 12-2 shows.
  • Page 149 Verification Set the DC Power Supply output voltage and CCH current as listed in Table 12-3 based on the test model. Press LOAD ON/OFF to start the Load and gradually reduce the DC Power Supply voltage till the DMM (V) shows the minimum operating voltage listed in Table 12-3 per model.
  • Page 150 High Power DC Electronic Load 63200 Series Operation & Programming Manual 0.939 A 1.061 A DMM Ai ± 0.061 10 A 9.98 A 10.02 A DMM Ai ± 0.02 63204_500V 0.1 A 0.0899 A 0.1101 A DMM Ai ± 0.0101 10 A 9.98 A...
  • Page 151: Cr Mode Verification

    Verification 12.3.2 CR Mode Verification This test module verifies if the programmed resistance complies with the specification in CR Mode. The calculation of programmed resistance is voltage divided current. The voltage (DMM (V)) passes through the input or measurement terminal of module. Also the voltage (DMM (I)) goes through the current shunt, that is shunt current = DMM (I) voltage/shunt resistance.
  • Page 152 High Power DC Electronic Load 63200 Series Operation & Programming Manual The current shunt range is 250A. Input the resistance listed in Table 12-6 and press LOAD ON/OFF to start the Load. Refer to the DMM (V) value and adjust the DC Power Supply value to be the same as the previous test model, and then wait for 30 seconds.
  • Page 153: Cv Mode Verification

    Verification The current shunt range is 250A. Input the resistance listed in Table 12-7 and press LOAD ON/OFF to start the Load. Refer to the DMM (V) value and adjust the DC Power Supply value to be the same as the previous test model, and then wait for 30 seconds. Record the DMM (V) voltage passed through the Load input terminal and the DMM (I) reading of shunt current.
  • Page 154 High Power DC Electronic Load 63200 Series Operation & Programming Manual Connect the DC Load, DC Power Supply, DMM and Current Shunt as Figure 12-1 shows. Use the DMM (V) to measure the voltage passing through the module input terminal.
  • Page 155 Verification Table 12-8 DMM(V) Model Front Panel Reading Voltage Setting Low Limit High Limit 70 V 69.885 V 70.115 V DMM Ai ± 0.075 63201 4.9175 V 5.0825 V DMM Ai ± 0.0425 400 V 399.3 V 400.7 V DMM Ai ± 0.45 63202_500V 10 V 9.495 V...
  • Page 156 High Power DC Electronic Load 63200 Series Operation & Programming Manual POINT VOLTAGE = Press ENTER again to go to the next screen and select 1. Von LATCH ENABLE 1: YES 2: NO Press ENTER again to go to the next screen and select the required current.
  • Page 157: Cp Mode Verification

    Verification 200 V 199.65 V 200.35 V DMM Ai ±0.225 63211 10 V 9.745 V 10.255 V DMM Ai ±0.13 200 V 199.65 V 200.35 V DMM Ai ±0.225 63212 10 V 9.745 V 10.255 V DMM Ai ±0.13 200 V 199.65 V 200.35 V DMM Ai ±0.225 63212...
  • Page 158 High Power DC Electronic Load 63200 Series Operation & Programming Manual 520 W 514.8 W 525.2 W DMM Ai ± 3.12 63204_600V 10 W 7.35 W 12.65 W DMM Ai ± 1.59 650 W 643.5 W 656.5 W DMM Ai ± 3.9...
  • Page 159: Dynamic & Slew Rate Circuit Testing

    Verification 6500 W 6435 W 6565 W DMM Ai ± 39 63205 650 W 614.25 W 685.75 W DMM Ai ± 21.45 10400 W 10296 W 10504 W DMM Ai ± 62.4 63206 1040 W 982.8 W 1097.2 W DMM Ai ± 34.32 10400 W 10296 W 10504 W DMM Ai ±...
  • Page 160 High Power DC Electronic Load 63200 Series Operation & Programming Manual 12.3.5.1 Checking for CC Dynamic Low Range First, press CC and DYNA., and then press RANGE to select CC Dynamic Low range. The LCD screen shows: CCDL1: 0.000A CCDL2: 0.000A...
  • Page 161 Verification 63205 6.75 6.75 63206 63207 6.25 6.25 63208 63209 63210_500V 63210_600V 63211 63212 PROBE OSCILLOSOPE AMPLIFIER DC POWER DC Load SUPPLY Figure 12-2 12-15...
  • Page 163: Appendix A List Of Functions & Mapped F/W Versions

    Additional functions and the mapped F/W version: 2.1 Add Timeout function in Battery 2.2 Modify the minimum setting of 63200 Series CV Mode to 0V 2.3 Add Recall Default function 2.4 External ON/OFF supports all Modes and Programs 2.5 Add SHORT KEY: ENABLE/DISABLE function 2.6 Change the text "CHROMA ATE Inc."...
  • Page 164 66 Huaya 1st Road, Guishan, Taoyuan 33383, Taiwan 台灣桃園市 33383 龜山區 華亞一路 66 號 T +886-3-327-9999 F +886-3-327-8898 Mail: info@chromaate.com http://www.chromaate.com Copyright by CHROMA ATE INC. All Rights Reserved. All other trade names referenced are the properties of their respective companies.