Query Example:
CONFigure:DIO:OUT2
Type:
Description:
Setting Syntax:
Setting Parameters: <NR1>, 0 ~ 7
Setting Example:
Query Syntax:
Return Parameters: <NR1>, 0 ~ 7
Query Example:
CONFigure:DIO:VOLTage
Type:
Description:
Setting Syntax:
Setting Parameters: <NR1>, 0 ~ 3
Setting Example:
Query Syntax:
Return Parameters: <NR1>, 0 ~ 3
Query Example:
DIO:OUT1
Type:
Description:
Setting Syntax:
Setting Parameters: <CRD | NR1>, OFF(0), ON(1)
Setting Example:
Query Syntax:
Return Parameters: <CRD>, OFF, ON
Query Example:
DIO:OUT2
Type:
Description:
CONF:DIO:OUT1?
Frame-Specific
Set the DO2 type the pin No.43 in System I/O Port.
CONFigure:DIO:OUT2<space><NR1>
0:NONE
1:OCP TEST PASS-H
2:OCP TEST FAIL-L
3:GONG TOTAL PASS-H
4:GONG TOTAL FAIL-L
5:OTP OVP OCP OPP REV-H
6:BUS CTRL. ACT-H
7:BUS CTRL. ACT-L
CONF:DIO:OUT2 1
CONF:DIO:OUT2 0
CONFigure:DIO:OUT2?
CONF:DIO:OUT2?
Frame-Specific
Set the voltage level of digital output pins (DO1&DO2).
CONFigure:DIO:VOLTage<space><NR1>
0:NONE
1:1.8V
2:3.3V
3:5.0V
CONF:DIO:VOLT 2
CONF:DIO:VOLT 3
CONFigure:DIO:VOLTage?
CONF:DIO:VOLT?
Frame-Specific
It sets t the system I/O port pin 9 DO1 status when the BUS CTRL.
mode is selected for DO1.
DIO:OUT1<space><NR1>
DIO:OUT1 ON
DIO:OUT1 0
DIO:OUT1?
DIO:OUT1?
Frame-Specific
It sets t the system I/O port pin 43 DO2 status when the BUS CTRL.
mode is selected for DO2.
Set DO2 to OCP TEST PASS-H.
Set DO2 to NONE.
[Unit = None]
Set output voltage is 3.3V.
Set output voltage is 5.0V.
[Unit = None]
Set DO1 to act.
Set DO1 not to act.
[Unit = None]