Cleaning and Disinfection
Wipe / Spray Disinfection

Preparing the Manual Wipe/Spray Disinfection

9-2
Heratherm
Disinfectants having an alcohol content of more than 10% may
form, in combination with air, easily combustible and explosive
gas mixtures.
When using such disinfectants, avoid open flames or exposure to
excessive heat during the entire disinfection process!
Use such disinfectants only in adequately ventilated rooms.
After the disinfectant has been allowed to react, wipe the cleaned
incubator components thoroughly dry.
Observe safety regulations to avoid fire and/or explosion hazard
caused by alcohol-containing disinfectants.
Chloride-containing disinfectants can corrode stainless steel.
Use only disinfectants that do not affect stainless steel!
The surfaces of the work space may be contaminated. Contact
with contaminated cleaning liquids may cause infections.
Disinfectants may contain harmful substances.
When cleaning and disinfecting, always observe the safety
instructions and hygiene guidelines!
• Wear safety gloves.
• Wear safety goggles.
• Wear mouth and respiratory system protection gear to protect your
mucous membranes.
• Observe the safety instructions of the disinfectant's manufacturer
and the hygiene supervisor.
Alcoholic disinfectants!
Chloride-containing disinfectants!
Health hazard