  • Page 1 Thermo Scientific Medilite Microcentrifuges Instruction Manual IM-448-3 June 2012 Visit us online to register your warranty. www.thermoscientific.com/labwarranty...
  • Page 2 © 2012 Thermo Fisher Scientific Inc. All rights reserved. Delrin, TEFLON, and Viton are registered trademarks of DuPont. Noryl is a registered trademark of SABIC. POLYCLEAR is a registered trademark of Hongye Chemical Co., Ltd. Hypaque is a registered trademark of Amersham Health As.
  • Page 3 WEEE Conformity This product is subject to the regulations of the EU Waste Electrical & Electronic Equipment (WEEE) Directive 2002/96. It is marked by the following symbol: Thermo Fisher Scientific has entered into agreements with recycling and disposal companies in all EU Member States for the recycling and disposal of this device.
  • Page 4: Table Of Contents

    Operation ..........................3-1 Warnings and Cautions ...................3-2 Operating Controls ....................3-2 Loading ........................3-2 Starting a Run ......................3-3 Stopping a Run .......................3-3 Chapter 4 Maintenance .......................... 4-1 Care and Cleaning ....................4-2 Chemical Compatibility Chart ................A-1 Contact Information ....................B-1 Thermo Scientific Medilite Microcentrifuges...
  • Page 5 Quantity Check Centrifuge Medilite Rotor IM448 CD with instruction manual If any parts are missing, please contact your nearest Thermo Fisher Scientific representative. This symbol refers to general hazards. WARNING means that injuries or material damage or contamination could occur.
  • Page 6 • Implementation of special measures which ensure that no one can approach the centrifuge for longer than absolutely necessary while it is running. • The mains plug must be freely accessible at all times. Pull out the power supply plug or disconnect the power supply in an emergency. Medilite Microcentrifuges Thermo Scientific...
  • Page 7 • Place the centrifuge on a clean, dry surface, to make certain that the suction feet at the bottom grip the surface firmly. • Especially when working with corrosive samples (salt solutions, acids, bases), the accessory parts and rotor chamber have to be cleaned carefully. • Always balance the samples. Thermo Scientific Medilite Microcentrifuges...
  • Page 8 Remaining risk: Inproper use can cause damages, contamincation, and inju- ries with fatal consequences. • Highly corrosive substances which can cause material damage and impair the mechanical stability of the rotor, should only be centrifuged in corresponding protective tubes. Medilite Microcentrifuges Thermo Scientific...
  • Page 9: Introduction

    Introduction Contents “Characteristics” page 1-2 • • “Technical Data” page 1-3 “Directives, Standards and Guidelines” page 1-4 • • “Mains Supply” page 1-4 “Rotor Selection” page 1-5 • Thermo Scientific Medilite Microcentrifuges...
  • Page 10: Characteristics

    • A brushless AC motor is attached to the cabinet. The motor is thermally protected in accordance with UL and CSA specifications. • A six or twelve place plastic 45° angle rotor. • A 30 minute timer (with hold function). • A viewport in the cover for easy speed verification. Medilite Microcentrifuges Thermo Scientific...
  • Page 11: Technical Data

    1   Introduction Technical Data Technical Data The technical data of the Thermo Scientific Medilite is listed in the following table. Table 1-1. Technical data Thermo Scientific Medilite Feature Value Environmental conditions -Indoor use only -Altitudes of up to 2,000 m above sea level -max.
  • Page 12: Directives, Standards And Guidelines

    Edition Mains Supply The following table contains an overview of the electrical connection data for the Thermo Scientific Medilite. This data is to be taken into consideration when selecting the mains connection socket. Table 1-3. Electrical connection data of theThermo Scientific Medilite...
  • Page 13: Rotor Selection

    1   Introduction Rotor Selection Rotor Selection The Thermo Scientific Medilite is supplied with a rotor. Two rotors are available to choose from. 6-Place Rotor 47430 12-Place Rotor 47620 Aeroshield 10 mL sealed with cap and o-ring 2087 Tube Size (mm) O.D.
  • Page 14: Installation

    Installation Contents “Receive the Unit ” page 2-2 • • “Prepare the Installation Site” page 2-2 “Verify Power Configuration ” page 2-2 • • “Moving the Unit ” page 2-3 Thermo Scientific Medilite Microcentrifuges...
  • Page 15: Receive The Unit

    Verify Power Configuration Verify that the correct power cord and connector is provided for your installation. The unit requires a grounded power supply (3-outlet). If your facility does not have grounded power outlets, arrange for proper grounding. Medilite Microcentrifuges Thermo Scientific...
  • Page 16: Moving The Unit

    4. Lift up an edge of the cup, and insert the flat object far enough to break the vacuum suction seal. 5. When all four suction cups are disengaged, lift the unit from the work surface. 6. When the unit is in its new location, ensure that the suction cups adhere correctly to the work surface. Thermo Scientific Medilite Microcentrifuges...
  • Page 17: Operation

    Operation Contents “Warnings and Cautions ” page 3-2 • • “Loading ” page 3-2 “Loading ” page 3-2 • • “Starting a Run ” page 3-3 “Stopping a Run ” page 3-3 • Thermo Scientific Medilite Microcentrifuges...
  • Page 18: Warnings And Cautions

    Mechanical damage can appear as cracks, scratches or gouges on the shield or rotor surfaces. Operating Controls Lid Opener Timer Loading 1. Unlatch and open the cover. 2. Check that the rotor turns freely and that there are no loose objects in the chamber. Medilite Microcentrifuges Thermo Scientific...
  • Page 19: Starting A Run

    Stopping a Run 1. To stop a run, turn the timer knob to zero or allow the time to expire. 2. Wait until the rotor has come to a complete stop, then unlatch and open the cover. Thermo Scientific Medilite Microcentrifuges...
  • Page 20: Maintenance

    Maintenance Contents “Care and Cleaning ” page 4-2 • Thermo Scientific Medilite Microcentrifuges...
  • Page 21: Care And Cleaning

    To remove the rotor, lift it straight up off its drive shaft. The rotor is held in place by spring clips that line up flat to the surface on the drive shaft. To reinstall the rotor, line up the clips with the flat surface and press it as far down the shaft as it will go. Medilite Microcentrifuges Thermo Scientific...
  • Page 22: Chemical Compatibility Chart

    U S S M S S U S S S S S S S S S U S S S S S S S S S S Cesium Acetate S S S - S S S S - S S - S S S S S S M S S S Thermo Scientific Medilite Microcentrifuges...
  • Page 23 U U U U S U U U S S M M S U U U - Hydrochloric Acid (conc.) U U U U - U U M - U M U U M U U U - U S U U U - Medilite Microcentrifuges Thermo Scientific...
  • Page 24 M S M - S S S S S S S M S M S S - N-Butyl Alcohol S U - S M - U S M S S S S M M S M - Thermo Scientific Medilite Microcentrifuges...
  • Page 25 S S S U S S S Nitric Acid (10%) U S U S S U U - S U S U - S S S S S S S S S M S S S S Medilite Microcentrifuges Thermo Scientific...
  • Page 26 Moderate attack, may be satisfactory for use in centrifuge depending on length of exposure, speed involved, etc. Suggest testing under actual conditions of use. Unsatisfactory, not recommended. Performance unknown; suggest testing, using sample to avoid loss of valuable material. Thermo Scientific Medilite Microcentrifuges...
  • Page 27 Chemical Compatibility Chart Chemical resistance data is included only as a guide to product use. No organized chemical resistance data exists for materials under the stress of centrifugation. When in doubt we recommend pretesting sample lots. Medilite Microcentrifuges Thermo Scientific...
  • Page 28: Contact Information

    +86 10 8419 3588 India 1800 22 8374 +91 22 6716 2200 Japan +81 45 453 9220 Other Asian Countries +852 2885 4613 Australia +61 39757 4300 New Zealand +64 9 980 6700 Other Countries +49 6184 90 6940 Thermo Scientific Medilite Microcentrifuges...