Figure 4-12 Connecting Nozzle for Fresh-air Filters (Accessory Item)

Access Port

A re-sealable, capped access port (can be closed off using the plugs delivered with the unit)
(Heratherm IMH 60/100/180 and IMH 60-S/100-S/180-S/400-S/750-S only, see figure 4-7)
allows cables, hoses or additional sensor leads to be routed into the work space of the
incubator.
The access port [2] has a diameter of 42 mm.
Shelf System
The incubator is supplied with two perforated shelves. The shelf support rails [1] have an
alternating pattern of oblong and round perforations spaced evenly at 30 mm, allowing the
shelf support brackets [8] to be inserted without any room for error, yet in a very flexible way
to accommodate any required height of sample container. The shelves [2] have an integrated
tilt protection and pull-out stop. For details on using the shelf system, see
System"
Thermo Scientific
When accessories are to be operated in the work space of the
incubator, the ambient condition requirements must be observed (see
table below). The energy introduced into the work space has an impact
on the lower end of the temperature control range. When additional
heating sources are introduced into the work space, temperature control
may be adversely affected.
on
page
5-1.
Work Space Components
Operating conditions
Product Description
"Installing the Shelf
Heratherm
4-15