Cleaning and Disinfection
Wipe / Spray Disinfection
Bio-hazard
Be sure to determine the current operating condition of the incubator
before you open the door!
Under specific circumstances, there is a risk that bacteria, viruses, fungi,
prions, and other biological substances survive when the
decontamination routine is aborted due to a power outage. Normally, the
incubator starts over with the decontamination cycle without any need
for user intervention upon return of power. Users who fail to notice the
power outage, may open the door and/or load the incubator with
samples, thus exposing themselves or their samples to bio-hazards
without being aware of that risk.
Decontamination must be performed immediately if any
biohazardous material is spilled in or on the incubator.