Thermo Scientific Heratherm IMH 60 Operating Instructions Manual page 113

Advanced protocol and advanced protocol security microbiological incubators
Hide thumbs Also See for Heratherm IMH 60:
Table of Contents

Advertisement

Table 13-1 Technical Data - Series IMH 60/100/180 and IMH 60-S/100-S/180-S
Parameter
Minimum bottom
clearance
Minimum top clearance
Table 13-2 Technical Data - Series IMH 400-S / IMH 750-S
Parameter
Process
Work space atmosphere,
Min.
Max.
Temperature deviation from set value at 37 °C
(99 °F), spatial.
Max. value/typical value
Temperature deviation from set value at 37 °C
(99 °F), over time
Heat-up time (work space unoccupied, from
25 °C (77 °F) to 98% of set temperature of
37 °C/99 °F))
Recovery time (work space unoccupied, door
open for 30 s.
Max. value/Typical value
Heat dissipation to environment (at set
temperature of 37 °C (99 °F) and room
temperature of 25 °C/77 °F)).
Overall dimensions
Height (with casters)
Width
Depth
Overall weight
Loading capacity
Max. overall loading capacity per device
Electrical data
Power rating
Max. Current
Earthing system (e. g. 1/N/PE)
Power line frequency
Thermo Scientific
Unit
IMH 60
IMH 100
mm/in
200/8
mm/in
300/12
Unit
°C/°F
°C/°F
K
K
min
min
W
mm/in
mm/in
mm/in
kg/lbs
kg/lbs
kg/lbs
W
A
Hz
IMH 180
IMH 60-S
IMH 100-S
IMH 400-S
Amb. temp. plus
5 °C/41°F
105 °C/221 °F
±0.3/±0.2
±0.2
35
<3/<2
87 ±10%
1655/
65.2
755/
29.7
770/
30.3
145/320
30/66
75/165
1380
11.5
1/N/PE
60
Technical Data
IMH 180-S
200/8
300/12
IMH 750-S
Amb. temp. plus
5 °C/41°F
105 °C/221 °F
±0.4/±0.3
±0.2
30
<3/<2
149 ±10%
1655/
65.2
1215/
47.8
770/
30.3
205/452
30/88
150/330
1560
13.0
1/N/PE
Heratherm
13-3

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents