Product Description
Work Space Components
Work Space Components
Inner Chamber
All components of the work space are made of corrosion-resistant stainless steel and have an
absolutely smooth and easy-to-clean surface. Any embossings have a large radius.
Connecting Nozzles for Fresh-air Filters (Accessory Items)
Observe the following note for IMH Series and IMH-S Series incubators with connecting
nozzles for fresh-air filters (accessory item).
A fresh-air filter can be attached to the connecting nozzle. After puncturing the plastic cap the
fresh-air filter can be screwed onto the nozzle.
The connecting nozzle can be closed off using the 1/8" closure plug delivered with the unit
after opening the plastic cap.
The connecting nozzle for the Heratherm IMH 60/100/180 and IMH 60-S/100-S/180-S
incubators has a G 1/8" threaded fitting.
4-14
Heratherm
Replacement should only be carried out by skilled and authorized
qualified personnel of electrotechnology/signal engineering!
The device fuses are not user-serviceable. When the incubator exhibits
the typical signs of a blown fuse (no response to pressing the On/Off
button, control panel remains extinguished, no heating operation), call
Thermo Fisher Scientific Customer Service to have the fuses replaced.
The connecting nozzle for the fresh-air filter (accessory item) may
neither be extended with nor be diverted through other tubing.
Fuse replacement
Connecting nozzle for fresh-air filters
Thermo Scientific