Operation
Starting Operation
5.
6. Load the work space with samples.
6-2
Heratherm
• refrain from loading the incubator with any of the substances
listed in the section
• make sure that the ambient air is free of any solvents
• do not operate the incubator in areas with an explosion hazard
The screen of the glass door, the interior panel of the outer door as
well as the surfaces of the shelving and the work space become hot
while the incubator is running through its heating cycles and need
some time to cool down.
When removing samples from a running or recently completed heating
cycle, always wear safety gloves and other appropriate personal
protection equipment to avoid burns on hot surfaces!
Overloading may damage the shelves or cause the shelves
and/or the incubator to tilt when the shelves are being drawn out,
ultimately destroying the samples. To avoid overloading the
incubator or its shelving be sure to observe the sample weight
limits specified in
chapter 13, "Technical Data."
To ensure sufficient air circulation and uniform heating of the samples,
do not use more than 70% of the maximum surface area of the work
space. Bulky objects in the work space that dissipate heat may impair
heat distribution.
To avoid any risk of explosion or fire
"Incorrect Use"
on
Hot surfaces
Risk of overloading
Proper loading
page 1-6