Handling and Control
Settings
Table 7-25 Entering the Calibration Reference Temperature Manually
2010-04-06
2010-04-06
2010-04-06

Date and Time

The Settings -> Time / Date option allows for switching between the international time and
date display formats and for setting the time and date of the internal clock. There are two
display formats to choose from:
• European date format DD.MM.YYYY and 24-hours time format. Example: 07.04.2010
and 15:05.
• US standard date format YYYY-MM-DD and 12-hour time format with AM/PM suffix.
Example: 2010-04-07 and 3:05 PM.
Table 7-26 Setting the Date Format
2010-04-06
7-24
Heratherm
1:33
Press
PM
Settings
confirm the selection with
Calibration
In the Calibration selection screen, press
1:33
PM
Settings
choose the preselected option Manual.
Calibration
Manual
In the settings dialog that appears, set the temperature
1:33
PM
measured with the external reference sensor by using
Settings
Calibration
36.9°C
The newly entered value will be stored and used to
calibrate the internal temperature sensors with the
value measured by the reference sensor.
The display returns to its default mode.
The Settings icon in the menu bar will go out.
Press
confirm.
The Error menu item from the Settings submenu
1:33
PM
appears in the multifunctional display pane.
Settings
Error
to switch to the Calibration menu item and
or
and confirm the settings with
to activate the menu bar, then use
to select the Settings icon and press
.
to
.
to