Fan

This menu item allows for turning on the fan inside the work space and gradually adjusting its
speed in one go. The current setting of the fan is indicated by the illuminated Fan icon in the
menu bar and the five-level bar graph (see item D6 in
setting located directly above the icon.
Table-top incubators
The fan can be adjusted in 6 levels:
• 0% (all chevrons extinguished)
• 20% (1 chevron illuminated)
• 40% (2 chevrons illuminated)
• 60% (3 chevrons illuminated)
• 80% (4 chevrons illuminated)
• 100% (5 chevrons illuminated)
Table 7-20 Turning on the Fan
31.03.2010
31.03.2010
31.03.2010
Thermo Scientific
Press
The settings dialog shown at left appears in the
14:15
multifunctional display pane now, with the value 0%
0%
flashing already.
Press
14:15
80%
speed level, then confirm with
The multifunctional display pane will show the current
fan speed level as a percentage (20%, 40%, 60%, 80%
14:15
or 100%). Additionally, the matching number of
100%
chevrons will be illuminated in the bar graph to the right.
The display returns to its default mode.
The Fan icon in the menu bar is illuminated now to
indicate that the fan is running.
figure 7-1
to activate the menu bar, then use
to select the Fan icon and press
as often as needed to reach the desired fan
Handling and Control
on
page
7-1) for the fan speed
or
to confirm.
.